Specifications - 2100 Bioanalyzer DNA Kits 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1

   

DNA 1000 Kit

DNA 7500 Kit

DNA 12000 Kit

High Sensitivity
DNA
Kit

Analytical Specifications

Sizing range

25–1000 bp

100–7500 bp

100–12000 bp

50-7000 bp

Sizing resolution

25–100 bp: 5 bp
100–500 bp: 5 %
500–1000 bp: 10 %

100–1000 bp: 5 %
1000–7500 bp: 15 %

100–1000 bp: 5 %
1000–12000 bp: 10 %

50-600 bp: ±10%
600-7000 bp:±20%

Sizing accuracy

+ 10 % CV*

+ 10 % CV*

+15 % CV*

± 10 % CV*

Sizing reproducibility

5% CV*

5% CV*

5% CV*

5% CV*

Quantitation accuracy

20% CV*

20% CV*

25% CV*

20% CV*

Quantitation reproducibility

25–500 bp:15 %CV*
500-1000 bp:5% CV*

100-1000 bp:10 % CV*
1000-7500 bp: 5 % CV*

100-1000 bp: 15 % CV*
1000-12000 bp: 10 % CV*

50-2000 bp: 15% CV*
2000-7000 bp: 10% CV*

Quantitative range

0.1 - 50 ng/µL*

0.1 - 50 ng/µL*

0.1 - 50 ng/µL*

5 -  500 pg/µL

Maximum salt concentration in sample

250 mM for KCl,
15 mM for MgCl2
250 mM NaCl

250 mM for KCl
15 mM for MgCl2
250 mM NaCl

250 mM for KCl
15 mM for MgCl2
250 mM NaCl

10 mM Tris and 1 mM EDTA

 

DNA 12/chip = 300 samples/kit

DNA 1000 Kit

DNA 7500 Kit

DNA 12000 Kit

High Sensitvity
DNA
Kit

Physical Specifications

Analysis time

35 minutes

30 minutes

30 minutes

45 minutes

Samples per chip

12

12

12

11

Sample volume

1 µL

1 µL

1 µL

1 µL

Kit stability

4 months (see box for storage temp.)

Kit size

25 chips

DNA 12/chip =
300 samples/kit

25 chips

DNA 12/chip =
300 samples/kit

25 chips

DNA 12/chip =
300 samples/kit

10 chips

DNA 11/chip =
110 samples/kit


* Respective DNA ladder as sample

 
 