Specifications 7100 Capillary Electrophoresis System

Pressure system
Programmable with -100 to +100 mbar on inlet
Flushing with 1 bar or with high pressure 2–12 bar
Vial pressurization with high pressure 2–12 bar on inlet and/or outlet

Safety features
Current leak detection  low current limit
Liquid leak sensor
Safety sensors at door and cover disabling high voltage
Vial sensor

Interfaces
LAN, CAN, RS232, remote control, analog in (1V, 20 bit, integrated A/D converter), analog out

Injection modes
Self correcting injection system with injection from inlet or outlet
Programmable range up to 10,000 seconds
Pressure       –100 to +100 mbar
Electrokinetic –30 to +30 kV

Autosampler/ fraction collector
50-position carousel
All vials are randomly accessible from inlet or outlet end of capillary
Temperature control with external waterbath with vial temperature from 10–40 °C
(non-condensing conditions, minimum waterbath temperature +1 °C)

Replenishment
Satellite station for refilling of inlet or outlet buffer vials with fresh buffer for automatic, continuous operation.
Selectable buffer levelling.

Vials
100 μL (polypropylene or glas)  with resealing snap caps
1 mL (polypropylene) with resealing snap caps
2 mL (glas) with resealing snap caps

Capillary cassette
High-speed forced-air cooler with Peltier element
Temperature range 10 °C below ambient, up to 60 °C  (min. 10 °C cassette temperature)
Minimum total capillary length 33 cm
Capillary compatibility 365 µm o.d.

Detector
Real time UV-Visible diode-array detector (190–600 nm)
Temperature controlled
Wavelength accuracy 1 nm
Response time 0.025 to 10 s
Light source prealigned deuterium or high brightness lamp
Signals up to eight signals simultaneously, full spectral acquisition with Agilent ChemStation

Sensitivity

1 μM 4-hydroxy-aceto-phenon injected at 50 mbar*5 sec, 3 x 50 µm bubble cell capillary, signal/noise >6* (50 mM borate buffer pH 9.3, 25 kV)
Baseline noise <50 μAU (at 2 sec response time)*
Linear dynamic range 1x104 (3x50 μm bubble cell capillary)*

Raw data channels
Detector signals and spectrum, voltage, current, leak current, power, cassette temperature, pressure, lamp voltage and detector temperature

Diagnostic functions
RFID tag for lamp, early maintenance feedback, supported by Agilent Lab Advisor software with integrated diagnosis test suite

System control
Operating with graphical user interface under Windows XP (SP3)
Time programmable parameters Time programmable parameters: voltage, current, power, polarity, pressure, inlet and outlet vial, capillary temperature, pre and post-run conditioning with pressure and/or voltage, Replenishment, fraction collection**

Electrophoresis power

Voltage range

setable 0 to ± 30 kV supply

Current

setable 0–300 μA

Power

setable 0–6 W

Operation under constant voltage, current or power

Programmable polarity switch

CE specific software
Mobility report, time corrected areas, pI calibration and bio polymer size calibration

Dimensions
Width 35 cm
Height 59 cm
Depth 51 cm
Weight 35 kg

Environment
Temperature 5–40 °C
Humidity up to 80 % relative humidity at 31 °C (non-condensing)

Power requirements
Line voltage 100-240 V, max. 300W
Line frequency 50-60 Hz

* typical value
**upgradeable

 
 