Trace Analysis of VOCs in City Air
Monitoring Air Pollutants with GC/MS and Canister Technology
by
P. Berno, V. Bonamin, and C. Gianoli,
SGS Ecologia, Padua (Italy)
Since the mid-eighties, several methods have emerged for
the sampling and analysis of VOCs
in both outdoor and indoor air based on the so-called canister
methodology.
The U.S. Environmental Protection Agency (U.S. EPA) introduced the most
complete analytical methods available to date: TO-14
(recently upgraded as TO-14A) and
IP-1A1,2 for the quantitative detection of
a subset of 41 of 187 hazardous air pollutants (HAPs) listed in the CAAA.
3
Subsequently, the ASTM published its own method suitable for "ambient, indoor,
or workplace atmospheres".4
More recently in Italy, an Environmental Ministry decree based on the canister
technology mandates benzene monitoring in urban
areas.5 A new method, known as TO-15,
was recently released to address a still larger number of VOCs, including polar organic
compounds, thereby covering a subgroup of 97 HAPs listed in the
CAAA.6 The TO-14 list,
incidentally, contains apolar (BTEX) or slightly polar compounds such as chloroethanes,
chloroethylenes, chlorobenzenes and various freons (see Figure 1).

Figure 1. Chromatogram of a TO-14 standard mixture.
- Freon 12
- Chloromethane
- Freon 114
- Vinyl chloride
- Bromo methane
- Ethyl Chloride
- Freon 11
- 1,1-dichloroethylene
- Methylene chloride
- Freon 113
- 1,1-dichloroethane
- 1,2-dichloroethylene
- Chloroform
- 1,2-dichloroethane
- Methylchloroform
- Benzene
- Carbon tetrachloride
- 1,2-dichloropropane
- Trichloroethylene
|
- cis- 1,3-dichloropropene
- trans- 1,3-dichloropropene
- 1,1,2-trichloroethane
- Toluene
- 1,2-dibromoethane
- Tetrachloroethylene
- Chlorobenzene
- Ethylbenzene
- m,p-xylene
- Styrene
- 1,1,2,2-tetrachloroethane
- o-xylene
- 1,3,5-trimethylbenzene
- 1,2,4-trimethylbenzene
- 1,3-dichlorobenzene
- 1,4-dichlorobenzene
- 1,2-dichlorobenzene
- 1,2,4-trichlorobenzene
- Hexachlorobutadiene
|
Canister Approach Superior
Because of the high dilution of the analytes of interest, both field sampling and analyses
must be carried out carefully. In our experience, the collection and analysis of samples
from canisters presents several advantages:
- Option of remote and unattended sampling
- Appropriate integration of the sample over a specific period of time
- Ease of storing and shipping, because the sample identity is unaffected for weeks
- Analysis of sample from multiple sites with a single analytical setup
- Repeatability of the analysis to allow, for example, assessment of measurement precision
The U.S. EPA Method TO-14 is composed of two parts: field sampling and lab analyses.
|
Field Sampling Using Canisters
Typical problems such as sample breakthrough or recovery efficiency that can be encountered
with the presently used solid sorbents, are totally eliminated. No sample
preparation is needed, which markedly shortens the analysis time.
Air samples are collected in stainless-steel containers known as canisters which underwent a
process of electro-passivation (SUMMAŽ process) to largely reduce the presence of polar active
sites on the inner surface. SilcoCans from Restek Corporation are even more inert, because
they have a chemically bonded fused-silica layer on the internal surface that further reduces
surface activity.
|
Table 1. Fields of application of the canister technique.
| Application |
Range |
| Urban Air Monitoring |
ppt-ppb |
| Landfill Site Gas Emissions |
ppb |
| Industrial Emissions |
ppb-ppm |
| Workplace Air Quality |
ppb-ppm |
| Indoor Air Monitoring |
ppt-ppm |
|
Eliminating Active Sites
The presence of active sites inevitably leads to undesired phenomena, such as irreversible
absorption of most pollutants of interest. Canisters are therefore cleaned before any sampling
campaign through a series of evacuation/filling cycles utilizing membrane and turbomolecular
pumps.
It is crucial to use oilless pumps and wet ultrapure nitrogen (whose function is
to saturate any residual active site still present on the surface with water molecules)
to achieve very low (sub-ppb)
levels of residual VOCs at the end of
the cleaning cycle.
It is strongly recommended to check
periodically for residual VOCs by performing blank runs after the cleaning process.
When the system is properly performing, this residual content should not exceed
0.01-0.1 ppb in each
component.
Controlling Passive Sampling
Passive sampling is carried out starting with a perfectly evacuated canister (internal pressure
less than 50 mtorr). Air is allowed to enter through a fine metering valve with a "critical
orifice" for careful control of the flow. The valve is set according to the desired sampling
time and/or canister volume using a mass
flow-meter and generally falls within the
3-100 sccm range. A mass flow meter is the most appropriate tool for a proper reading of
the flow. With a 6-liter canister, sampling may last from a few minutes
to 24 hours. When longer sampling times are required, larger canisters (up to 33 liters) or
active sampling (canister
pressurization) may be used.
Variety of City Pollutants
Table 2 shows our results for a replicate
measurement of a sample of urban air collected in daytime
and close to the rush hour. As expected, the sample contains a variety of hydrocarbons from car
exhausts. Traces of other pollutants such as freons and halogenated solvents have been detected
at ppb-level.
The results shown are part of a 12-day benzene monitoring carried out in Padua in
collaboration with the local environmental authorities. The study was aimed
at providing a verification of the data obtained from the air monitoring network
located throughout the city along main roads and intersections.
We collected two samples a day, the first around the rush hour from 8 a.m. to 9 a.m.,
and the second averaged over 24 hours. The results reflect a single-point sample
(one-hour sampling) collected twice
independently. The very good reproducibility reflects optimum sample stability vs. time
and the fine control of
the analytical conditions the system allows. The system easily meets the most demanding
standards imposed by several local and international regulations aimed at improving air quality.
|
Standards and Equipment
Certified gaseous calibration
standard mixtures of these
compounds, normally at 1-ppm
levels and pressurized in
electro-polished aluminum
cylinders, are commercially
available. The analytical equipment (see photo) is
composed of an Entech Model 7000 preconcentrator,
which acts as an injector to the GC/MS (HP 6890/5972).
Other indispensable
tools are the canister cleaner (Entech 3000SL) and the
dynamic diluter (Entech 4650SL) for standards preparation.
|
| Analyte |
Variation (x)/X % |
Mean Conc. (x) (pg/m3) |
Conc. x1 (pg/m3) |
Conc. x2 (pg/m3) |
| 1. Freon 12 (dichlorodifluoromethane) |
7.4 |
2.7 |
2.6 |
2.8 |
| 2. Methyl chloride |
0.0 |
1.4 |
1.4 |
1.4 |
| 3. Freon 114 (1,2-dichlorotetrafluoroethane) |
|
|
ND |
ND |
| 4. Vinyl chloride |
|
|
ND |
ND |
| 5. Methyl bromide |
|
|
ND |
ND |
| 6. Ethyl chloride |
|
|
ND |
ND |
| 7. Freon 11 (trichloromethane) |
3.6 |
2.8 |
2.8 |
2.7 |
| 8. 1,1-dichloroethane |
|
|
ND |
ND |
| 9. Methylene chloride |
4.7 |
4.3 |
4.4 |
4.2 |
| 10. Freon, 113 (1,1,2-trichlorotrifluoroethane) |
11.8 |
0.9 |
0.8 |
0.9 |
| 11. 1,1-dichloroethane |
|
|
ND |
ND |
| 12. cis-1,2-dichloroethane |
|
|
ND |
ND |
| 13. Chloroform |
0.0 |
0.2 |
0.2 |
0.2 |
| 14. 1,2-dichloroethane |
|
|
ND |
ND |
| 15. 1,1,1-trichloroethane |
0.0 |
1.0 |
1.0 |
1.0 |
| 16. Benzene |
1.9 |
15.7 |
15.5 |
15.8 |
| 17. Carbon tetrachloride |
0.0 |
0.9 |
0.9 |
0.9 |
| 18. 1,2-dichloropropane |
0.0 |
0.6 |
0.6 |
0.6 |
| 19. Trichloroethylene |
0.0 |
0.9 |
0.9 |
0.9 |
| 20. cis- 1,3-dichloropropene |
|
|
ND |
ND |
| 21. trans- 1,3-dichloropropene |
|
|
ND |
ND |
| 22. 1,1,2-trichloroethane |
|
|
ND |
ND |
| 23. Toluene |
3.4 |
47.0 |
46.2 |
47.8 |
| 24. 1,2-dibromoethane |
|
|
ND |
ND |
| 25. Tetrachloroethylene |
0.0 |
1.5 |
1.5 |
1.5 |
| 26. Chlorobenzene |
|
|
ND |
ND |
| 27. Ethylbenzene |
0.0 |
10.1 |
10.1 |
10.1 |
| 28. m,p-xylene |
0.9 |
35.0 |
35.1 |
34.8 |
| 29. Styrene |
0.0 |
2.4 |
2.4 |
2.4 |
| 30. 1,1,2,2- tetrachloroethane |
|
|
ND |
ND |
| 31. o-xylene |
0.0 |
13.0 |
13.0 |
13.0 |
| 32. 1,3,5-trimethylbenzene |
0.0 |
4.8 |
4.8 |
4.8 |
| 33. 1,2,4-trimethylbenzene |
0.6 |
16.1 |
16.0 |
16.1 |
| 34. m-dichlorobenzene |
0.0 |
0.1 |
0.1 |
0.1 |
| 35. p-dichlorobenzene |
0.0 |
0.5 |
0.5 |
0.5 |
| 36. o-dichlorobenzene |
|
|
ND |
ND |
| 37. 1,2,4-trichlorobenzene |
0.0 |
0.1 |
0.1 |
0.1 |
| 38. Hexachloro-1,3-butadiene |
0.0 |
0.1 |
0.1 |
0.1 |
| |
| Sum of VOCs |
1.1 |
161.9 |
161.0 |
162.7 |
| ND = Not Detected |
Table 2. VOCs measured in the city of Padua during the rush hour. Two independent samples
of urban air were collected in parallel.
| Conditions |
|
Preconcentrator
|
| Configuration: |
Cold Trap Dehydration |
| Carrier: |
He |
| |
Module 1 |
Module 2 |
Module 3 |
| Trapping Vol. (mL) |
|
400 |
|
| Trapping Flow (mL/min) |
|
60 |
|
| Trapping Temp (°C) |
-10 |
-70 |
-150 |
| Trans. Temp (°C) |
20 |
190 |
70 |
| Trans. Flow (mL/min) |
10 |
1.5 |
1.5 |
| Trans. Vol. (mL) |
10 |
|
|
| Trans. Time (min) |
|
3.5 |
2 |
| Gas Chromatograph |
| Column: |
HP-1; methylsilicone; 60 m X 0.32 mm X 1.0 µm film |
| Carrier: |
He |
| Flow: |
1.5 mL /min (constant) |
| Injector Temp: |
250°C |
| Transfer Line: |
280°C |
| Oven: |
35°C X 10 min
to 150°C at 7°C/min; to 220°C at 20°C/min X 6 min |
| Mass Spectrometer |
| Start Time: |
3.8 min |
| Mass Range: |
33.0 to 270 amu |
| Threshold: |
100 |
| Sampling: |
2 |
| Scans/sec: |
3.13 |
The Analytical Process
The system is designed so that sample
transfer from the canister to the GC column
is quantitative, accurate and reproducible.
The larger the sample volume, the more critical
to eliminate most of the water and carbon dioxide
in the steps preceding
the injection.
The diagram shows the Entech three-module
water management system. In the Microscale
purge-and-trap configuration, 20 to 1000 milliliters
of sample are transferred from the canister to a glass
beads trap (Module 1), kept at -150°C with liquid nitrogen.
Here VOCs are quantitatively blocked together with most of
the carbon dioxide and water.
Module 1 is then warmed up to room temperature,
and volatiles are transferred by a He stream to a
Tenax trap kept at -10°C. At this temperature, most
of the CO2 is not retained, while VOCs remain adsorbed
on the Tenax packing of Module 2. Desorption takes place by heating
Module 2 to 180°C, which allows sample cryofocusing in Module 3
to be cooled to liquid-nitrogen temperature.
After rapid injection and separation on a
high-resolution methylsilicone column (HP-1),
VOCs are analyzed by a quadrupolar mass-selective
detector (see chromatographic conditions).
Data are processed with an HP-Vectra PC using
the SmartLabTM package
developed at Entech for the
control of the Model 7000 preconcentrator, the 3000SL
canister cleaner and the 4650SL dynamic diluter, while
the HP 6890 Series MSD software controls the GC/MS.
Quantitation can be done using appropriate standards (either
internal or external) or even labelled standards when very
high precision is required.
|
SGS Ecologia in Brief
SGS Ecologia srl is part of the environmental branch of the
SGS Group based in Geneva (Switzerland). Founded in 1878, the
Group is the world's largest organization in the field of inspection
and verification. Truly worldwide and benefiting from a unique
international network of affiliated companies with over 1,180 offices,
321 laboratories and 33,000 employees, the SGS Group is able to provide
its international clientele with a comprehensive range of services
in more than 140 countries.
SGS Ecologia assists customers and experts in the field
providing air quality assessments, indoor and outdoor ambient
air monitoring, industrial-gas emission testing and monitoring
services, and a wide range of other ecological-assessment
services.
|
REFERENCES:
- U.S. EPA Compendium Method TO-14A,
Determination of Volatile Organic Compounds
(VOCs) in Ambient Air Using Specially Prepared
Canisters with Subsequent Analysis by Gas Chromatography (1997).
- U.S. EPA Method IP-1A, Determination
of Volatile Organic Compounds (VOCs) in
Indoor Air (1989).
- Clear Air Act Amendments of 1990, United
States Code, Title III - Hazardous Air Pollutants.
- ASTM Method D5466-93, Standard Test
Method for Determination of Volatile Organic
Chemicals in Atmospheres (Canisters Sampling
Methodology), Annual Book of ASTM Standards,
Vol. 11.03, p. 404 (1995).
- D.M. n. 159, 25 Nov. 1994; G.U. Suppl. Ord.
n. 290 del 13 Dic. (1994).
- U.S. EPA Compendium Method TO-15,
Determination of Volatile Oraganic Compounds
(VOCs) in Air Collected in Specially Prepared
Canisters and Analyzed by Gas Chromatography/
Mass Spectrometry (GC/MS) (1997).
ACKNOWLEDGEMENT
We thank Dr. L. Baracco of the Ufficio Tutela Ambiente,
Provincia di Padova for permission to reproduce some
of the data obtained during the 1996 VOC monitoring in
the city of Padua.
|