|
QUALIFICATION/VERIFICATION OF INSTALLATION, OPERATION AND PERFORMANCE
Concerned With Compliance? We Can Help!

IQ and OQ/PV services include
comprehensive handbooks to guide the instrument user. |
In many business fields, analysts must comply with a number of regulations. Although the ultimate
responsibility for complicance remains with the user, the manufacturers of analytical equipment can
assist in a variety of ways. |
Among the pertinent regulations are Good Laboratory Practice (GLP)
and Good Manufacturing Practice (GMP), and quality or accreditation standards such as ISO 9000,
European Norm EN 45001, and ISO Guide 25. Pharmaceutical analysts must consider the requirements of
the United States and European Pharmacopeias (USP and EP).
Hewlett-Packard offers services for
installation qualification (IQ) and operational qualification/ performance verification (OQ/PV) for
its major products in the capillary electrophoresis (CE), high-
performance liquid chromatography
(HPLC) and UV-visible spectroscopy product lines.
Users Determine Limits
Installation qualification is included in the standard installation service and can be requested by users. IQ verifies and documents
completeness of shipment. Usually, a single measurement is performed to prove system functionality.
For CE and HPLC products, a standard sample is injected; for UV-visible spectroscopy products,
a spectrum is measured. Standards and procedures are included with the shipment, if required.
Operational qualification/performance verification is designed to verify and
document the
instrument's ability to meet certain performance criteria. Users determine acceptance limits according
to their own specific needs. In general, OQ/PV performs several system tests, reports the results
and documents whether the acceptance criteria were met.
Qualifying an HPLC System
To qualify a complete HPLC system,
a series of tests must be performed. These tests include noise, drift,
linearity and wavelength accuracy of the detector; carryover, linearity and reproducibility of
the injector; gradient performance of the pump; and temperature accuracy and precision of the column
compartment. The user must define acceptance criteria for each.
Once scheduled, the tests can be performed automatically with the HP ChemStation
for HPLC, using a set of
calibrated and traceable standards for most. The ChemStation
then calculates and prints a report in which the actual results are compared with specified
acceptance criteria.
Testing a CE System
The OQ/PV test for the HP 3DCE system should
be performed before its first use in regulated environments, after extended use and, in particular,
after repair. Procedures used to test the instrument's functionality have been developed specifically
for capillary electrophoresis.
The OQ/PV kit includes a users
guide, test sample, buffer and
capillary, and a 3.5-inch diskette containing the methods, sequences and spectral library
necessary for running the test. The users guide describes how to prepare the system, how to
run each test, and how to interpret the results. It includes advice on appropriate action
should the system fail the test. The users guide is written in the form of a standard operating procedure
(SOP); the individual methods and sequences are also included. The entire test procedure is
designed to be performed in half a day.
OQ/PV for UV-Visible Systems
The HP ChemStation software for the HP 8453 UV-visible spectroscopy system includes a semiautomatic procedure for OQ/PV of the spectrophotometer,
but obtaining suitable standards remains difficult. Solid standards, such as those available from NIST, are expensive and
do not fully meet the regulatory requirements. For example, a good, solid standard for the photometric accuracy
in the UV region is not available. Liquid standards meet regulatory needs, but need to be freshly prepared before
the test,
are time-consuming and prone to error.
To solve these problems, Hewlett-Packard has introduced a
set of low-cost, prepared liquid standards that are easy to use. Simply break open the ampoule, pour into the
cuvette and measure. The only additional requirement is water as a blank
for the noncritical stray-light tests.
All other reference solvents are included.
Three stray-light tests are included to
meet both European
Pharmacopeia and U.S. regulatory needs. The new standards are easy to use and meet the regulatory requirements exactly.
They come complete with a certificate of analysis for traceability.
With the liquid standards, flow-through
cuvettes are recommended to improve sample handling. The HP-developed procedure uses a syringe to introduce the sample,
significantly reducing test time and the risk of cross-contamination.
The procedures for performing OQ/PV
are fully documented.
IQ and OQ/PV Service
HP offers IQ and OQ/PV services. Specially trained service personnel
perform all procedures using calibrated tools and standards. The services save valuable time and allow analysts to
concentrate on more important tasks.
|