Early Diagnosis of Transformer Troubles


Early Diagnosis of Transformer Troubles

Automated Dissolved-Gas Analysis Of Transformer Oils

by Jocelyn Jalbert and Roland Gilbert, Institut de Recherche en Electricité Hydro­Québec (Canada) and Susan Brillante, Hewlett­Packard Co.

The malfunctions of high­voltage power transformers can be diagnosed by analyzing the dissolved gases that typically form in the insulating materials of the devices. A method for automated transformer­oil gas analysis (TOGA) has been developed by combining an HP 7694 static headspace sampler with an HP 5890 Series II capillary gas chromatograph and an HP 3365 ChemStation.

Background of TOGA

High­voltage power transformers are susceptible to problems such as arcing, overheating and partial discharges. These malfunctions always result in the chemical decomposition of the insulating materials, mineral oil and cellulose. Decomposition, in turn, produces several gases that are totally or partially dissolved in the oil, including H2, CO, CO2, CH4, C2H2, C2H4, C2H6 and C3H8. The contact of ambient air with the oil may also cause gases such as O2, N2 and CO2 to be present.

Because the relationship of dissolved gases to specific transformer problems are known, it is possible to diagnose the presence of faults at an early stage and take preventive action before failures (which can include explosion) occur. Early diagnosis of faults by performing TOGA on a routine basis will result in cost savings and improvements in service reliability.

TCD and FID Detectors

An HP 7694 headspace sampler (HS) equipped with a six­port valve and a 1­mL sample loop was used to reproducibly sample the dissolved gases in transformer oils. The method uses two detectors in series: a thermal conductivity detector (TCD) for the quantitation of H2, O2 and N2 and a flame ionization detector (FID) with a methanizer for the quantitation of the other gases.

The gases were separated by using a molecular­sieve PLOT column connected to a Carboxen(TM) PLOT column through a 6­port valve. The TCD and FID signals were recorded simultaneously by the ChemStation. The figures show the FID signal and the TCD signal, respectively, for a primary oil sample.

Standard Method Allows No Automation

Most electrical utilities perform TOGA according to Method A of the D­3612 standard procedure approved by the American Society for Testing and Materials (ASTM). This method uses vacuum extraction of the gases of interest, followed by packed­column GC chromatography. The method cannot be automated and, to achieve reliable results, requires a dedicated and well­trained analyst.

The authors have analyzed transformer oil samples using both the ASTM D­3612 Method A procedure and the HS/capillary GC technique (Jalbert & Gilbert, 1996). The HS results demonstrated excellent correlation with the ASTM method and lower detection limits.

Why Headspace?

Compared with the ASTM method, HS/capillary GC offers significant benefits:
  • equivalent precision
  • lower detection limits
  • shorter analysis times
  • automation

These benefits result in higher sample throughput in the analytical laboratory. The analysis takes 55 minutes for the first sample, and 25 minutes for each additional one. It is at least five times faster than the ASTM D­3612 method and potentially yields higher­quality results with less skilled operators.

Reference

Jalbert, J., and Gilbert, R. Decomposition of Transformer Oils: A New Approach for the Determination of Dissolved Gases. Accepted for presentation to the IEEE Power Engineering Society, Denver, CO (1996).