GC/MS HELPS STOP DRUG ABUSE BY PLAYERS
World Cup Kicks Out Drugs
Performance-enhancing drugs continue to threaten the integrity
of sporting events and the health and well-being of participants.
Drug testing therefore has become a routine practice at major
sporting events such as the 1998 FIFA Soccer World Cup (WC '98).
This year, the World Cup involves more teams than ever, and the
organizers are committed to catching any players misusing drugs.
The Laboratoire National de Despistage du Dopage in Paris,
France, one of only 23 laboratories in the world accredited by
the International Olympic Committee (IOC), will handle all drug
testing for WC '98. HP analytical instruments such as the HP
GC/MS system will play a critical role in the analysis of urine
samples from players participating in the event.
250-300 Banned Substances
Each urine sample sent to the laboratory must be tested
for 250-300 different substances and metabolites banned by FIFA --
no small task even with today's highly sophisticated analytical
instruments. Banned substances fall into five main categories:
- Stimulants. This category includes common drugs such as
caffeine as well as cocaine and amphetamines
- Analgesics. Morphine and other narcotics allow athletes
to continue to train or compete when injured, often causing
serious damage and posing the threat of addiction.
- Anabolic Steroids. These muscle-building drugs include
naturally occurring hormones, such as testosterone and their
derivatives.
- Diuretics. They cause increased urine excretion and
are used to mask the use of other banned substances.
- Peptide Hormones. Possibly the most difficult group
to detect reliably, they include growth hormone and corticotrophin,
which is used to increase the levels of corticosteroids and
induce euphoria.
Quantitation Important for Some
In most cases, tests for these substances are qualitative.
Any use of a banned substance, in any quantity, contravenes FIFA's rules,
and even tiny traces of a banned substance or its metabolites
indicate a positive result. Some of the banned compounds occur
in the body naturally, such as testosterone, while others are
prohibited only in quantities considered abusive, caffeine, for
example. In these cases, the substances must be quantitated, and
only levels above the designated threshold are considered positive.
Consistent Data Handling Essential
Athletes' careers and reputations depend on the laboratory
results; standard operating procedures are therefore put in
place for each type of testing. The laboratories operate under
Good Laboratory Practice (GLP) and regularly undergo a stringent
review by the IOC.
The Paris laboratory has a series of HP GC/MS systems that are
linked by an HP ChemLAN laboratory network -- ensuring that methods
and analytical data are handled and recorded consistently. In the
event of a legal challenge, the system provides evidence that the
correct operation has taken place and eliminates the possibility
of sample misidentification.
The Techniques Used
Good sample preparation is essential. Five
different procedures are used, which include passing the sample
through columns that bind specific types of compounds. Contaminating
substances can be washed away and molecules of interest concentrated.
The concentrated sample undergoes a number of additional steps,
including hydrolysis, extraction, and concentration, depending on
which substances it is being tested for. Sample preparation involves
a mixture of manual and automated processes and takes about two hours.
Gas chromatography combined with
mass spectroscopy is the main technique used by the laboratory
to detect banned substances. Together, GC and MS provide a "fingerprint"
for each compound. The laboratory checks the mass spectra for compounds
with molecular weights characteristic of banned substances.
Then, at the touch of a button, the laboratory can check the retention time on
the GC. Only if both retention time and mass are consistent with a banned
substance or a known metabolite is the sample considered positive.
Analgesics, anabolic steroids, stimulants, and diuretics can all be
identified using this method. GC/MS is also used to quantitate
levels of testosterone, caffeine, and ephedrine.
Banned Drugs Detectable 99%
To detect and quantitate corticosteroids, the laboratories
use HPLC in conjunction with mass spectroscopy. HPLC is also
used to test for the presence of diuretics, while immunoassays
are used to detect peptide hormones. Current analytical techniques
can detect 99% of the compounds banned by FIFA.
To ensure that compounds are not misidentified, standards
and controls are essential. Developing a standard for only the
banned substances would not be difficult for the laboratory -- a
standard sample of the drug could be added to a urine sample, for
example. But laboratories must be able to detect all metabolites
of a banned substance, and these can be assessed accurately only
by letting a volunteer take the banned drug and using the urine
as the positive standard. The negative standard is the urine from
a volunteer who has not taken any banned substances. All urine
samples tested must be run with the appropriate positive and
negative standards.
Quality and accuracy requirements are extremely strict, and
laboratories regularly receive faked samples to test their procedures.
A single false positive result is sufficient for a laboratory to
lose its IOC accreditation. In the event of a false negative result,
the accreditation will be suspended until the reason is found and
corrective action has been taken.
HP makes the ideal partner for such drug monitoring missions;
first by offering a complete range of high-quality analytical
instrumentation, then by providing the necessary applications
and maintenance support.
|