Question: Should I just select the Bonus-RP column for all my method
development with basic compounds?
Answer: Basic compounds are best analyzed following the method development
scheme outlined earlier in this brochure. Select an
SB-C8 or
SB-C18 column for initial
development and use a buffered low pH mobile phase. Many times this approach provides a
good separation and the StableBond columns will have exceptional lifetime at low pH even
at high temperatures.
Increasing Retention of Basic Compounds
Question: I cant get enough retention of my basic compounds at low pH,
what column should I try next?
Answer: The
Eclipse XDB column
can be used in the intermediate pH region
from pH 3 - 8 for the longest lifetime. The primary advantage of using the intermediate pH
region is a possible increase in retention for basic compounds. Most basic compounds have
pKa values of 5 or greater, therefore at pH 6 - 8 some of these compounds may become
non-charged and the column will retain them more. In addition, at pH > 5 the residual
silanols on the silica surface will become charged. This can lead to stronger interactions
with basic compounds and increase retention. The Eclipse XDB column is the ideal column to
use in this mid-pH region because the dense-bonding and double endcapping will cover the
most active silanols on the surface of the column, and any residual silanols can
contribute to increased retention of basic analytes without causing excessive peak
tailing. Figure 1 shows a plot of retention vs pH for a group of basic compounds. At pH 6
and higher the retention of all of these compounds increases due to increased interaction
with the column, though at pH 6.5 only Triprolidine is non-charged.
If retention has not increased enough, then the next step would be to try the
Extend-C18. All of the basic compounds shown in Figure 1, except Triprolidine, have pKa
values of 9.0 - 9.2. Therefore they must be analyzed at pH 10 or higher before they are
non-charged. The Extend-C18 can effectively be used at this high pH to improve the
retention of basic compounds.
FIGURE 1

Effect of pH on Retention Antihistamines
Column: Eclipse XDB-C8 (Agilent Part No. 993967-902)
Mobile Phase: 75% 25 mM Phosphate buffer
25% Acetonitrile
Temperature: 40°C
Back to Top
Choosing Bonded-Phases
Question: What column do you recommend to analyze raw materials?
Answer:There is not enough information provided in this question to give a specific recommendation. Column choice, and appropriate mobile phase, depends on many characteristics of the sample such as polarity, pKa of ionizable functional groups, solubility vs. pH, and molecular weight.
Question: Im currently using a C18 column and my separation has a couple
of peaks that elute early and a couple of peaks that elute late. What can I do to reduce
the analysis time and maintain resolution of the early eluting peaks?
Answer: First, you could try a gradient elution method on the C18 column. But
many people do not like to use gradients, so choosing a different bonded-phase may help.
Short chain polar bonded-phases such as the
SB-CN and
SB-Phenyl are ideal for separations
like this. The increased polarity of these phases reduces retention of the later eluting
hydrophobic compounds while often maintaining the retention of earlier eluting hydrophilic
compounds. Figure 2 shows this clearly. Using the same mobile phase conditions these three
compounds are well resolved on the SB-CN in about 5 minutes. The same separation on the
SB-C8 column takes nearly twice as long and provides incomplete resolution.
FIGURE 2

SB-C8 (Agilent Part No. 866953-906)
SB-CN (Agilent Part No. 866953-905)
Separation of Acid/Base Sample on StableBond
Columns: 4.6 x 75 mm, 3.5 µm
Mobile Phase: 80% 25 mM NaH2PO4, pH 3.0 20% MeOH
Flow Rate: 1.0 mL/min
Temperature: 35°C
Detection: UV 254 nm
Sample:
1. Barbital
2. Sulfamethoxazole
3. Caffeine
Question: When should I select Phenyl bonded-phases?
Answer: Both
SB-Phenyl
and
Eclipse XDB-Phenyl bonded phases provide unique
selectivity and are often a good choice for changing the selectivity for two closely
eluting analytes. Phenyl bonded phases are less retentive than C8 bonded-phases so they
offer a second option for reducing retention of late eluting hydrophobic compounds and
minimizing analysis time (Figure 3).
FIGURE 3
SB-C18 (Agilent Part No. 863953-902)

SB-Phenyl (Agilent Part No. 863953-912)
Distinct Selectivity of ZORBAX SB-Phenyl and SB-C18 Bonded Phases
Columns: 4.6 x 150 mm, 3.5 µm
Mobile Phase: 84% MeOH 16% 0.1% TFA H2O, pH 2.0
Flow Rate: 1.0 mL/min
Temperature: 30°C
Detection: UV 310 nm
Sample:
1. Oxybenzone
2. Padimate-O
3. Ethylhexyl-p-methoxycinnamate
4. Ethylhexyl-salicylate
Back to Top
Temperature Control
Question
: What temperature should I use for my separation?
Answer: Temperature control for separations is important for long-term
retention reproducibility, one factor of method ruggedness. Controlling temperature at 35
- 40°C is normally sufficient for good method reproducibility and ruggedness. In
addition, the use of elevated temperature can have other benefits. First, it reduces the
system operating pressure by reducing the viscosity of the mobile phase.
Second, it will reduce analysis time, which can substantially increase productivity
(Figure 4).
Third, temperature may change the selectivity of a separation. Not all compounds have
the same response to temperature so the selectivity of a separation can change
dramatically when temperature is increased or decreased.
StableBond columns have high
temperature limits
SB-C18 can be taken up to 90°C, at low pH, and
SB-C8
,
SB-Phenyl
,
SB-CN
and
SB-C3
can be taken up to 80°C, making it possible to optimize your
separation without changing columns, particularly if you are analyzing ionizable
compounds.
FIGURE 4

High Temperature Can be Used with StableBond Columns for High Throughput Analysis
Columns: 4.6 x 75 mm, 3.5 µm
ZORBAX Rapid Resolution SB-C18 (Agilent Part No. 866953-902)
Mobile Phase: 26% MeOH
74% H2O + 10 mM Hexane Sulfonate and 0.1% phospheric acid
Sample: Water Soluble Vitamins
Back to Top
Custom Columns
Question
: There are a lot of different column configurations (dimensions)
available, but I dont see the one Im looking for. Can I get a column made in
the configuration I want?
Answer: Most likely. Agilent is continually adding to its HPLC column
offerings, so check with your
Agilent column distributor or in the US call Agilent
Customer Assist at 800-227-9700 and ask for HPLC column support to find out if the column
configuration you need is currently available. For locations outside the U.S., your
authorized Agilent column distributor, listed on the Agilent website, can also help you
find out the cost and delivery time. If the column is not available, a column can be
packed in the configuration of your choice with available bonded-phases. Select an
internal diameter and specify a column length, bonded-phase, and particle size using
product number 899999-999. A special products quote (SPQ) will then be issued to track
your order.
Rapid Resolution Columns
Question: I typically select 4.6 x 250 mm, 5 µm columns for my analytical work
because I have complex samples, but I need to reduce my analysis time and increase my
sample throughput. What can I do?
Answer:
Rapid Resolution
columns are ideal for your needs. A Rapid Resolution
4.6 x 150 mm, 3.5 µm column will reduce your analysis time by 40% while maintaining your
resolution. Your gains are in reduced analysis time, whether you are doing isocratic or
gradient separations, and substantial solvent savings (Figure 5). You may be able to
choose even shorter Rapid Resolution column lengths and maintain the desired resolution
(Rs) because Rs a N1/2. This means that decreasing column length, and therefore
efficiency, will not decrease Rs by the same amount. It is likely that resolution will be
maintained on even shorter column lengths and choosing 75 mm or even 50 mm column lengths
can reduce analysis time even more.
FIGURE 5

SB-C8 (Agilent Part No. 883975-906)
SB-C8 (Agilent Part No. 866953-906)
Rapid Resolution Chromatography
Column: 4.6 x 150 mm, 5 µm
ZORBAX StableBond SB-C8 (Agilent Part No. 883975-906)
Gradient Time: 30 min.
Flow Rate: 1.0 mL/min
Analysis Time: 24 min.
Column: 4.6 x 150 mm, 5 µm
Column: 4.6 x 75 mm, 5 µm
ZORBAX StableBond SB-C8 (Agilent Part No. 866953-906)
Gradient Time: 15 min.
Flow Rate: 1.0 mL/min
Analysis Time: 12 min.
Sample:
1. Tebuthiuron
2. Prometon
3. Prometryne
4. Atrazine
5. Bentazon
6. Propazine
7. Propanil
8. Metochlor
Question: But are these columns more likely to fail than 5 µm columns?
Answer: No, they are not. The
Rapid Resolution
columns are as rugged as the 5
µm, 250 mm columns. There are two types of accelerated column failure attributed to using
smaller particles. First, smaller particle columns are thought to plug faster. This is not
true when a standard 2 µm frit is used at the top of the column. Because of careful
particle size control and the use of 3.5 µm particles, ZORBAX Rapid Resolution columns
will not contain any particles as small as 2 µm. This means a standard frit can be used
on the column and the Rapid Resolution columns will be no more prone to plug than a 5 µm
particle size column.
Second, columns with smaller particles are thought to have shorter lifetimes because
the column beds compress, leaving voids that cause peak broadening and tailing. It is true
that 3.5 µm particle size columns will operate at slightly higher pressures than 5 µm
columns, but ZORBAX particles can easily withstand these increases in pressure. ZORBAX
particles are packed at 8000 psi and can easily withstand pressures up to 5000 psi in
routine use. A Rapid Resolution 4.6 x 150, 3.5 µm column will typically be operated below
3000 psi, so the column bed will not compress when using ZORBAX Rapid Resolution columns.
So both the rugged ZORBAX particles and the standard 2 µm column frit assure you long
column lifetime when using Rapid Resolution columns.
FIGURE 6

Run Time: 1.8 min.
Equilibration Time: 1 min.
Total Analysis Time: 2.8 min.
Fast Gradient Analysis of Heart Drugs on a Very Short Column
Column: 4.6 x 50 mm, 3.5 µm
Rapid Resolution ZORBAX Eclipse XDB-C8 (Agilent Part No. 935967-906)
Mobile Phase: A: 25 mM Na2HPO4, pH 3.0
Gradient: 42 - 74% B in 1.8 min.
Flow Rate: 3.0 mL/min
Temperature: 35°C
Sample: Cardiac Drugs
1. Diltiazem
2. Dipyridamole
3. Nifedipine
4. Lidoflazine
5. Flunarizine
The very short Rapid Resolution columns can also be used at high flow rates to further
reduce analysis time. Figure 6 shows a very rapid gradient analysis at 3 mL/min on a 4.6 x
50 mm Rapid Resolution Eclipse XDB-C8 column.
Question
: What do you think about neutralizing extra-column volume by packing short columns with wide-bores (7mm)?
Answer: As long as you are willing to accept operating at a higher flow rate. If you operate at 1.0 mL/min on a 4.6 mm i.d.
column, then operating at 2.3 mL/min would be the equivalent linear velocity when using a 7.0 mm i.d. column. Solvent waste may be an issue.
High-Throughput HPLC/Combinatorial Chemistry
Question: My work requires very high sample throughput. How fast can I do a
gradient separation?
Answer: More and more people, especially those who analyze combinatorial
chemistry samples, need to analyze a lot of unknown samples quickly. The best way to do
this is on short,
Rapid Resolution
columns with rapid gradient times. The best gradient
time is the one that resolves all of your analytes in the least amount of time. On very
short columns 30 and 50 mm lengths a good starting point would be a 2 - 5
minute gradient. From there optimize your separation for organic range and gradient time.
On these short columns you can easily increase the flow rate to further decrease analysis
time to 30 seconds without exceeding the pressure limits of the columns. Column
re-equilibration times are typically as short as the analysis times 3 - 5 minutes
for the 50 mm columns.
Question: Can I really effectively use these very short columns on my HPLC
instrument?
Answer: Yes, you can. For columns of 3.0 mm i.d. and above, no instrument
adjustments are necessary. For gradient separations with 2.1 mm and 1.0 mm i.d. columns
the ideal HPLC is a high-pressure mixing instrument like the
Agilent 1100 HPLC
with
the binary pump, because it minimizes the gradient delay volume. Using a low volume mixer
and the injector by-pass (or micro-injector) further minimize gradient delay volume.
Narrow i.d. tubing and a low volume detector flow cell are preferred but not necessary.
These changes are easy to make and allow you to effectively use columns as small as the
2.1 x 30 mm columns or even the 2.1 x 15 mm columns (Figure 7).
FIGURE 7

SB-C18 (Agilent Part No. 873700-902) SB-C18 (Agilent Part No. 875700-902)
Gradient High-Throughput Narrow-Bore HPLC . . in Less Than 60 Seconds
Column: 2.1 x 30 mm, 3.5 µm
Rapid Resolution StableBond SB-C18 (Agilent Part No. 873700-902)
Gradient: 5 - 25% B in 1.0 min.
Flow Rate: 1.0 mL/min
Analysis Time : 100 sec.
Column: 2.1 x 15 mm, 3.5 µm
Rapid Resolution StableBond SB-C18 (Agilent Part No. 875700-902)
Gradient: 10 - 60% B in 30 sec.
Flow Rate: 2.0 mL/min
Analysis Time : 47 sec.
Mobile Phase: A: Water with 0.2% formic acid
B: Methanol with 0.2% formic acid
Temperature: 35°C
Sample: Organic Acids
1. Gallic
2. Protocatechuric
3. Hydrocaffeic
4. Gentisic
5. Vanillic
6. Syringic
7. Sinapinic
8. Salicyclic
Back to Top
LC/MS
Question: Very short narrow-bore columns seem ideal for LC/MS. What bonded
phase should I use?
Answer: LC/MS requires volatile mobile phases and the most suitable buffers are
acetate and formate. The buffer range for acetate is from pH 3.8 - 5.8 so the best
bonded-phases to start with would be the
Eclipse XDB C18 and
C8. Formate is also widely
used and has a buffer range from pH 2.8 - 4.8. When using this buffer you can select
StableBond
columns then
Eclipse XDB
columns and follow the same method development scheme
as discussed in the method development section. For high pH, using ammonium hydroxide at
pH 10.5, select Extend-C18.
Answer: Detergents are not recommended for cleaning reversed-phase (RP) HPLC columns. Ionic detergents are typically long-chain
carbon compounds having an ionizable group at one end and have been used as ion-pairing agents. Because these long carbon chains partition
well into the bonded phase of reversed-phase columns (i.e. the bonded-phase of the C18 reversed-phase column strongly retains the long carbon
chain of a detergent), the removal of detergents can be difficult, if not impossible.
Answer: HPLC columns are more efficient and packed better than years ago, therefore this is generally recommended for most reversed-phase and normal phase silica based columns. This procedure is designed to remove particles from the column frit when high pressure occurs at the column inlet, but it will not work all the time. Because it does not require opening the column it is worth trying. Make sure when you do this that you disconnect the column from the detector and make sure the particles that plugged the column are not coming from the HPLC system or you may just plug the frit at the back end of the column and this may not be replaceable. Columns can also be turned in the reverse direction for washing/cleaning with stronger solvents to remove adsorbed material. This has the benefit of not exposing the rest of the column to the contaminants. When this is done the column should also not be attached to the detector.
Answer: Some detergents are used as ion-pair (IP) reagents, e.g. SDS with a carbon chain length of twelve. However, most commonly
used ion-pair reagents have shorter carbon chains, e.g. hexane sulfonate has carbon chain length of six. IP reagents with longer carbon chains are more
difficult to remove from RP-HPLC columns. Studies have shown that ion-pair reagents and some detergents are best removed with long washes with a
50/50: v/v, methanol/water mobile phase system. If you can regain the retention, selectivity and efficiency (resolution) of a known separation
(e.g., QC sample) after washing, then you might have successfully removed the ion-pair reagent. I would not recommend using this column for developing
a new method as exposure to ion-pair reagents may change the retention characteristics of a column permanently. If scouting runs for a new method
are conducted on a column exposed to ion-pair reagents, a new column should be purchased as soon as possible to verify that the selectivity is not different.
In my opinion, columns exposed to IP reagents should be dedicated to the ion-pair method.
Answer: Much as you would any RP-phase column and I've included some general instructions provided that you are using standard mobile
phase conditions to separate small molecules. If you are separating peptides and proteins or if your sample is dissolved in plasma, then the guidance is
different and I recommend that you call 800-227-9770, press option 1 and ask for HPLC column technical support for more information (within the US, other
areas contact your local Agilent sales office.
Answer: A phosphoric acid wash has been shown to be effective at reducing tailing caused by the sample complexing with metals in the HPLC
system. Typically a 1% phosphoric acid wash of the system and column is suggested to eliminate this tailing and it works. It is perfectly reasonable to use
these recommended wash conditions with Agilent ZORBAX StableBond reversed-phase HPLC products. The StableBond HPLC column is particularly stable
at low pH - the SB-C18 column is stable at a pH of 0.8 and 90°C. How can you tell if your peak tailing is caused by metal complexation? Look to see if a
lone pair of electrons on either a N or O atom can chelate with the metal to form a 5 or 6-membered ring. Metal complexation is a commonly overlooked
cause of peak tailing and metals are presents in every HPLC system.
Answer: I assume you are using a reversed-phase column. Please review the recommended
HPLC column cleaning procedure
. I
do suggest that if column performance declined quickly that you consider using a
guard column
and possibly consider including a preliminary clean-up
step before injection.
Answer: Wash the column in the reverse direction, not attached to the detector and at half the recommended flow rate (keep the pressure
below the recommended maximum). First choose a solvent that will dissolve what you believe has contaminated the column. Most GPC columns are
PS-DVB and you need to check the solvent compatibility before using a solvent. Many wash solvents are a higher viscosity than the typical eluting
solvents so a lower flow rate with careful attention to pressure is needed.
Anionic samples can adsorb onto PS-DVB and if these have
contaminated your GPC column a wash solution with a salt is recommended. Check to see what types of salts are recommended for the column. In
some cases the polarity of the material adsorbed may require washing with organic solvents modified with acid (formic or acetic) or base (triethanolamine)
(check the pH range) or some water may be compatible with an appropriate organic solvent. If more hydrophobic material were retained, elevated temperature
along with an appropriate organic solvent would be recommended. Once again you need to check the maximum temperature range allowed for your column.
If you wash carefully the column should not degrade from the solvent switching.
But there are at least four other causes of peak fronting. These are column channeling,
ionic interactions between the analytes and the silica, poor solubility of the analyte in
the mobile phase, and wettability problems between the mobile phase and bonded-phase. A
problem with column channeling should result in peak fronting on all peaks or at least the
largest peak in the chromatogram if the others are very small. If this occurs you need to
replace the column. To evaluate this problem try a new column.
Increasing the buffer ionic strength or changing the pH of the mobile phase can often
improve ionic interactions, which cause fronting. Increasing the ionic strength can reduce
interactions between the silica and ionic analytes and changing the pH can have the same
effect.
Solubility needs to be assessed by trying to improve the solubility of the analyte and
evaluating the resulting chromatogram. For instance, you can increase the time you
sonicate your sample and re-inject or dissolve it in a solvent where the sample has good
solubility and then dilute and inject in the mobile phase. Also any sample you suspect has
a solubility problem should be filtered. You can also mix your sample and mobile phase
off-line to see if it is visibly soluble in the mobile phase.
Wettability refers to the ability of the mobile phase to fully penetrate the
bonded-phase so that analytes interact with all of the bonded-phase. In cases where there
is a highly aqueous mobile phase with a C18 column complete wettability may not be
achieved. The bonded-phase can fold over on itself. The results may be loss of retention
and distortions in peak shape. If you can, increase the amount of organic in the mobile
phase and re-evaluate peak shape. If not you may need to consider selecting a column
designed for use in very high aqueous mobile phases.
These are the typical causes of peak fronting and some solutions. Please note that some
of these problems may also result in peak tailing, though the solutions would still be the
same.