<%flg="B"%>

GC Troubleshooting

 

 

Evaluating The Problem
The first step in any troubleshooting effort is to step back and evaluate the situation. Rushing to solve the problem often results in a critical piece of important information being overlooked or neglected. In addition to the problem, look for any other changes or differences in the chromatogram. Many problems are accompanied by other symptoms. Retention time shifts, altered baseline noise or drift, or peak shape changes are only a few of the other clues that often point to or narrow the list of possible causes. Finally, make note of any changes or differences involving the sample. Solvents, vials, pipettes, storage conditions, sample age, extraction or preparation techniques, or any other factor influencing the sample environment can be responsible.

Back to Top

 

Checking The Obvious
A surprising number of problems involve fairly simple and often overlooked components of the GC system or analysis. Many of these items are transparent in the daily operation of the GC and are often taken for granted (set it and forget it). The areas and items to check include:

  • Gases - pressures, carrier gas average linear velocity, and flow rates (detector, split vent, septum purge).
  • Temperatures - column, injector, detector and transfer lines.
  • System parameters - purge activation times, detector attenuation and range, mass ranges, etc.
  • Gas lines and traps - cleanliness, leaks, expiration.
  • Injector consumables - septa, liners, O-rings and ferrules.
  • Sample integrity - concentration, degradation, solvent, storage.
  • Syringes - handling technique, leaks, needle sharpness, cleanliness.
  • Data system - settings and connections.

Back to Top

 

Ghost Peaks or Carryover
System contamination is responsible for most ghost peaks or carryover problems. If the extra ghost peaks are similar in width to the sample peaks (with similar retention times), the contaminants were most likely introduced into the column at the same time as the sample. The extra compounds may be present in the injector (i.e., contamination) or in the sample itself. Impurities in solvents, vials, caps and syringes are only some of the possible sources. Injecting sample and solvent blanks may help to find possible sources of the contaminants. If the ghost peaks are much broader than the sample peaks, the contaminants were most likely already in the column when the injection was made. These compounds were still in the column when a previous GC run was terminated. They elute during a later run and are often very broad. Sometimes numerous ghost peaks from multiple injections overlap and elute as a hump or blob. This often takes on the appearance of baseline drift or wander.

Increasing the final temperature or time in the temperature program is one method to minimize or eliminate a ghost peak problem. Alternatively, a short bake-out after each run or series of runs may remove the highly retained compounds from the column before they cause a problem. Performing a condensation test is a good method to determine whether a contaminated injector is the source of the carryover or ghost peaks.

Back to Top

 

Excessive Baseline Noise

Possible Cause Solution Comments
Injector contamination Clean the injector Try a condensation
test; gas lines may
also need cleaning
Column contamination Bake-out the column Limit the bake-out
to 1-2 hours
Column contamination Solvent rinse the column Only for bonded
and corss-linked
phases
Detector contamination Clean the detector Usually the noise increases
over time and not suddenly
Contaminated or low
quality gases
Use better grade gases;
also check for expired
gas traps or leaks
Usually occurs after
changing a gas cylinder
Column inserted too far
into detector
Reinstall the column Consult GC manual for
the proper insertion distance
Incorrect detector gas flow rates Adjust the flow rates to the
recommended values
Consult GC manual for
the proper flow rates
Leak when using an MS,
ECD or TCD
Find and eliminate the leak Usually at the column
fittings or injector
Old detector filament,
lamp or electron multiplier
Replace appropriate part  

Back to Top

 

Baseline Instability or Disturbances

Possible Cause Solution Comments
Injector contamination Clean the injector Try a condensation test; gas lines may also need cleaning
Column contamination Bake-out the column Limit bake-out to 1-2 hours
Unequilibrated detector Allow the detector to stabilized Since detectors may require up to 24 hours to fully stabilize
Incompletely conditioned column Fully condition the column More critical for trace level analysis
Change in carrier gas flow rate during the temperature program Normal in many cases MS, TCD and ECD respond to
changes in carrier gas flow rate

Back to Top

 

Tailing Peaks

Possible Cause Solution Comments
Column Contamination Trim the column Remove 1/2-1 meter from the front of the column
Column Contamination Solvent rinse the column Only for bonded and cross-linked phases
Column activity irreversible Only affects active compounds
Solvent-phase polarity mismatch Change sample solvent More tailing for the early eluting peaks or those closest to solvent front
Solvent-phase polarity mismatch Install a retention gap 3-5 meter retention gap is sufficient
Solvent effect violation for splitless or on-column injections Decrease the initial column tempterature Peak tailing decreases with retention
Too low of a split ratio Increase the split ratio Flow from split vent should be 20 mL/min or higher
Poor column installation Reinstall the column More tailing for the early eluting peaks
Some active compounds always tail None Most common for amines and carboxylic acids

Back to Top

 

Split Peaks

Possible Cause Solution Comments
Injection technique Change technique Usually related to erratic plunger depression or  having sample in the syringe needle
Mixed sample solvent Change the sample solvent to a single  solvent Worse for solvents with large differences in polarity or boiling points
Poor column
installation
Reinstall the column in the injector Usually a large error in the insertion distance
Sample degradation in the injector Reduce the injector temperature Peak broadening or tailing may occur if the temperature is too low
Sample degradation in the injector Change to an on-column injector Requires an on-column injector

Back to Top

 

Retention Time

Possible Cause Solution Comments
Change in carrier gas velocity Check the carrier gas velocity All peaks will shift in the same direction by approximately the same amount
Change in column temperature Check the column temperature Not all peaks will shift by the same amount
Change in column dimension Verify column identity  
Large change in compound concentration Try a different sample concentration May also affect adjacent peaks
Leak in the injector Leak check the injector A change in peak size also usually occurs.
Blockage in a gas line Clean or replace the plugged line More common for the split line; also check flow controllers and solenoids

Back to Top

 

Change in Peak Size

Possible Cause Solution Comments
Change in detector response Check gas flows, temperatures and settings All peaks may not be equally affected
Change in detector response Check background level or noise May be caused by system contamination and not the detector
Change in the split ratio Check split ratio All peaks will not by equally affected
Change in the purge activation time Check the purge activation time For splitless injectors
Change in injector volume Check the injection technique Injection volumes are not linear
Change in sample concentration Check and verify sample concentration Changes may also be caused by degradation, evaporation, or variances in sample temperature or pH
Leak in the syringe Use a different syringe Sample leaks passed the plunger or around the needle; leaks are often not readily visible
Column contamination Trim the column Remove 1/2-1 meter from the front of the column
Column contamination Solvent rinse the column Only for bonded and cross-linked phases
Column activity Irreversible Only affects active compounds

Back to Top

 

Loss of Resolution

Possible Cause Solution Comments
Decrease in Separation
Different column temperature Check column temperature Differences in other peaks will be visible
Different column dimensions or phase Verify column identity Differences in other peaks will be visible
Coelution with other peak Change the column temperature Decrease column temperature and check for the appearance of a peak shoulder or tail
Increase in peak width
Change in carrier gas velocity Check carrier gas velocity A change in retention time also occurs
Column contamination Trim the column Remove 1/2 to 1 meter from the front of the column
Column contamination Solvent rinse the column Only for bonded and cross-linked phases
Column contamination Trim the column Remove 1/2-1 meter from the front of the column
Column contamination Solvent rinse the column Only for bonded and cross-linked phases
Change in the injector Check the injector settings Typical areas: split ratio, liner, temperature, injection volume
Change in sample concentration or solvent Try a different sample concentration Peak widths increase at higher concentrations

Back to Top

 

Condensation Test
Use this test whenever injector or carrier gas contamination problems are suspected (e.g., ghost peaks or erratic baselines).

  • Leave the GC at 40-50°C for 8 or more hours.
  • Run a blank analysis (i.e., start the GC, but with no injection) using the normal temperature conditions and instrument settings.
  • Collect the chromatogram for this blank run.
  • Immediately repeat the blank run as soon as the first one is completed. Do not allow more than 5 minutes to elapse before starting the second blank run.
  • Collect the chromatogram for the second blank run and compare it to the first chromatogram.
  • If the FIRST chromatogram contains a substantially larger amount of peaks and baseline instability, then that is an indication that there is contamination upstream of the capillary column (ie. contaminated inlet, dirty carrier gas, etc.).
  • If BOTH chromatograms contain few peaks or very little baseline drift, it can be assumed that the carrier gas and/or inlet are relatively clean.
  • If BOTH chromatograms contain a significant amount of noise and/or baseline drift, then that usually is an indication that the detector or detector gases are contaminated.

 

.

Back to Top

 

"Low-Bleed" Columns - Fact or Fiction?
Prof. Walt Jennings
Cofounder, J&W Scientific Incorporated

Several manufacturers offer "low bleed" columns. In some cases, these are merely selected from the standard production process, but in other cases the columns are actually "synthesized" for low bleed. In recent years, it has been established that where functional groups (i.e. phenyl) are inserted into the polysiloxane chain as aryl inclusions, as opposed to being attached to the chain as pendant groups, the resultant phase possesses increased thermal and oxidative resistance. Columns coated with such phases emit lower levels of bleed signal and are capable of going to higher temperatures. The increased thermal resistance is apparent only at temperatures above ca. 300 degrees. While some users can reap the benefits of these developments, others find little or no improvement.. their bleed signals are still too high.

True column bleed, of course, comes only from the column. What the user perceives as bleed is usually the total signal reaching the detector, which is the summation of signal from the septum (this gives a typical silicone mass spectrum), the injector, and the detector, all of which is usually blamed on the column.

It is good procedure to first check the detector. Disconnect and remove the column, and place an undrilled cap on the column attachment fitting. Activate the detector, and note the signal at 50 degrees. Increase the oven temperature to 320 degrees, and again note the signal. On a pristine detector, the FID signal will increase by one to two picoamps. If the increase exceeds this level, attention should be directed to cleaning the detector, make-up gas and hydrogen lines. Once the detector signal falls to an acceptable level at 320 degrees, attention should be directed to the injector. If the injector liner is visibly soiled, the injector should be cooled, dissembled and interior cavities scrubbed with solvent and natural bristle brushes or cotton swabs. After assembling the injector, a "jumper tube" (one to three meters of uncoated fused silica or steel tubing) is then used to connect the injector directly to the detector. The injector heater should be energized, and the oven set at 320 degrees. Any increase in "bleed" signal over that observed with the detector alone must come from the front end of the instrument, and may originate with the septum, the carrier gas line, in-line regulators, valves, or flow controllers.

Wrap a new septum in aluminum foil, ensuring that one face is smooth, and install this, smooth side down. If the signal emanating from the jumper tube is decreased, it indicates a need for better quality septa. If the signal is still high, materials entrained in the carrier gas may have deposited in lines, valves, or regulators, which should be dissembled and cleaned or replaced.

When the combined signal from the injector and detector falls to an acceptable level (one to two picoamps @ 320 degrees on an FID), the user is ready to install and reap the benefits of a true low-bleed column. The bleed rate of conventional columns is normally high enough to mask signal from the injector and detector unless these latter are heavily contaminated. With low bleed columns, the signal from the injector and detector assumes increased importance. This spurious signal is not infrequently limiting, and is usually (and incorrectly) perceived as "column bleed".

Back to Top