|
GC Troubleshooting
Evaluating
The Problem
The first step in any troubleshooting effort is to step back and
evaluate the situation. Rushing to solve the problem often results in a
critical piece of important information being overlooked or neglected.
In addition to the problem, look for any other changes or differences in
the chromatogram. Many problems are accompanied by other symptoms.
Retention time shifts, altered baseline noise or drift, or peak shape
changes are only a few of the other clues that often point to or narrow
the list of possible causes. Finally, make note of any changes or
differences involving the sample. Solvents, vials, pipettes, storage
conditions, sample age, extraction or preparation techniques, or any
other factor influencing the sample environment can be responsible.
Back
to Top
Checking
The Obvious
A surprising number of problems involve fairly simple and often
overlooked components of the GC system or analysis. Many of these items
are transparent in the daily operation of the GC and are often taken for
granted (set it and forget it). The areas and items to check include:
- Gases - pressures, carrier gas average
linear velocity, and flow rates (detector, split vent, septum
purge).
- Temperatures - column, injector,
detector and transfer lines.
- System parameters - purge activation
times, detector attenuation and range, mass ranges, etc.
- Gas lines and traps - cleanliness,
leaks, expiration.
- Injector consumables - septa, liners,
O-rings and ferrules.
- Sample integrity - concentration,
degradation, solvent, storage.
- Syringes - handling technique, leaks,
needle sharpness, cleanliness.
- Data system - settings and
connections.
Back
to Top
Ghost
Peaks or Carryover
System contamination is responsible for most ghost peaks or carryover
problems. If the extra ghost peaks are similar in width to the sample
peaks (with similar retention times), the contaminants were most likely
introduced into the column at the same time as the sample. The extra
compounds may be present in the injector (i.e., contamination) or in the
sample itself. Impurities in solvents, vials, caps and syringes are only
some of the possible sources. Injecting sample and solvent blanks may
help to find possible sources of the contaminants. If the ghost peaks
are much broader than the sample peaks, the contaminants were most
likely already in the column when the injection was made. These
compounds were still in the column when a previous GC run was
terminated. They elute during a later run and are often very broad.
Sometimes numerous ghost peaks from multiple injections overlap and
elute as a hump or blob. This often takes on the appearance of baseline
drift or wander.
Increasing the final temperature or time
in the temperature program is one method to minimize or eliminate a
ghost peak problem. Alternatively, a short bake-out after each run or
series of runs may remove the highly retained compounds from the column
before they cause a problem. Performing a condensation test is a good
method to determine whether a contaminated injector is the source of the
carryover or ghost peaks.
Back
to Top
Excessive
Baseline Noise
| Possible
Cause |
Solution |
Comments |
| Injector
contamination |
Clean the
injector |
Try a
condensation
test; gas lines may
also need cleaning |
| Column
contamination |
Bake-out the
column |
Limit the
bake-out
to 1-2 hours |
| Column
contamination |
Solvent rinse
the column |
Only for bonded
and corss-linked
phases |
| Detector
contamination |
Clean the
detector |
Usually the
noise increases
over time and not suddenly |
Contaminated or
low
quality gases |
Use better grade
gases;
also check for expired
gas traps or leaks |
Usually occurs
after
changing a gas cylinder |
Column inserted
too far
into detector |
Reinstall the
column |
Consult GC
manual for
the proper insertion distance |
| Incorrect
detector gas flow rates |
Adjust the flow
rates to the
recommended values |
Consult GC
manual for
the proper flow rates |
Leak when using
an MS,
ECD or TCD |
Find and
eliminate the leak |
Usually at the
column
fittings or injector |
Old detector filament,
lamp or electron
multiplier |
Replace
appropriate part |
|
Back
to Top
Baseline
Instability or Disturbances
| Possible
Cause |
Solution |
Comments |
| Injector
contamination |
Clean
the injector |
Try
a condensation test; gas lines may also need cleaning |
| Column
contamination |
Bake-out
the column |
Limit
bake-out to 1-2 hours |
| Unequilibrated
detector |
Allow
the detector to stabilized |
Since
detectors may require up to 24 hours to fully stabilize |
| Incompletely
conditioned column |
Fully
condition the column |
More
critical for trace level analysis |
| Change
in carrier gas flow rate during the temperature program |
Normal
in many cases |
MS,
TCD and ECD respond to
changes in carrier gas flow rate |
Back
to Top
Tailing
Peaks
| Possible
Cause |
Solution |
Comments |
| Column
Contamination |
Trim the column |
Remove
1/2-1 meter from the front of the column |
| Column
Contamination |
Solvent rinse
the column |
Only for bonded
and cross-linked phases |
| Column activity |
irreversible |
Only affects
active compounds |
| Solvent-phase
polarity mismatch |
Change sample
solvent |
More tailing for
the early eluting peaks or those closest to solvent front |
| Solvent-phase
polarity mismatch |
Install a
retention gap |
3-5 meter
retention gap is sufficient |
| Solvent effect
violation for splitless or on-column injections |
Decrease the
initial column tempterature |
Peak tailing
decreases with retention |
| Too low of a
split ratio |
Increase the
split ratio |
Flow from split
vent should be 20 mL/min or higher |
| Poor column
installation |
Reinstall the
column |
More tailing for
the early eluting peaks |
| Some active
compounds always tail |
None |
Most common for
amines and carboxylic acids |
Back to
Top
Split
Peaks
| Possible
Cause |
Solution |
Comments |
| Injection
technique |
Change
technique |
Usually
related to erratic plunger depression or having sample
in the syringe needle |
| Mixed
sample solvent |
Change
the sample solvent to a single solvent |
Worse
for solvents with large differences in polarity or boiling
points |
Poor
column
installation |
Reinstall
the column in the injector |
Usually
a large error in the insertion distance |
| Sample
degradation in the injector |
Reduce
the injector temperature |
Peak
broadening or tailing may occur if the temperature is too low |
| Sample
degradation in the injector |
Change
to an on-column injector |
Requires
an on-column
injector |
Back to
Top
Retention
Time
| Possible
Cause |
Solution |
Comments |
| Change in
carrier gas velocity |
Check the
carrier gas velocity |
All peaks will
shift in the same direction by approximately the same amount |
| Change in column
temperature |
Check the column
temperature |
Not all peaks
will shift by the same amount |
| Change in column
dimension |
Verify column
identity |
|
| Large change in
compound concentration |
Try a different
sample concentration |
May also affect
adjacent peaks |
| Leak in the
injector |
Leak check the
injector |
A change in peak
size also usually occurs. |
| Blockage in a
gas line |
Clean or replace
the plugged line |
More common for
the split line; also check flow controllers and solenoids |
Back to
Top
Change
in Peak Size
| Possible
Cause |
Solution |
Comments |
| Change in
detector response |
Check gas flows,
temperatures and settings |
All peaks may
not be equally affected |
| Change in
detector response |
Check background
level or noise |
May be caused by
system contamination and not the detector |
| Change in the
split ratio |
Check split
ratio |
All peaks will
not by equally affected |
| Change in the
purge activation time |
Check the purge
activation time |
For splitless
injectors |
| Change in
injector volume |
Check the
injection technique |
Injection
volumes are not linear |
| Change in sample
concentration |
Check and verify
sample concentration |
Changes may also
be caused by degradation, evaporation, or variances in sample
temperature or pH |
| Leak in the
syringe |
Use a different syringe |
Sample leaks
passed the plunger or around the needle; leaks are often not
readily visible |
| Column
contamination |
Trim the column |
Remove
1/2-1 meter from the front of the column |
| Column
contamination |
Solvent rinse
the column |
Only for bonded
and cross-linked phases |
| Column activity |
Irreversible |
Only affects
active compounds |
Back to
Top
Loss
of Resolution
| Possible
Cause |
Solution |
Comments |
| Decrease
in Separation |
| Different column
temperature |
Check column
temperature |
Differences in
other peaks will be visible |
| Different column
dimensions or phase |
Verify column
identity |
Differences in
other peaks will be visible |
| Coelution with
other peak |
Change the
column temperature |
Decrease column
temperature and check for the appearance of a peak shoulder or
tail |
| Increase
in peak width |
| Change in
carrier gas velocity |
Check carrier
gas velocity |
A change in
retention time also occurs |
| Column
contamination |
Trim the column |
Remove
1/2 to 1 meter from the front of the column |
| Column
contamination |
Solvent rinse
the column |
Only for bonded
and cross-linked phases |
| Column
contamination |
Trim the column |
Remove
1/2-1 meter from the front of the column |
| Column
contamination |
Solvent rinse
the column |
Only for bonded
and cross-linked phases |
| Change in the
injector |
Check the
injector settings |
Typical areas:
split ratio, liner, temperature, injection volume |
| Change in sample
concentration or solvent |
Try a different
sample concentration |
Peak widths
increase at higher concentrations |
Back to
Top
Condensation
Test
Use this test whenever
injector or carrier gas contamination problems are suspected (e.g.,
ghost peaks or erratic baselines).
- Leave the GC at 40-50°C for 8 or more
hours.
- Run a blank analysis (i.e., start the
GC, but with no injection) using the normal temperature conditions
and instrument settings.
- Collect the chromatogram for this
blank run.
- Immediately repeat the blank run as
soon as the first one is completed. Do not allow more than 5 minutes
to elapse before starting the second blank run.
- Collect the chromatogram for the
second blank run and compare it to the first chromatogram.
- If the FIRST chromatogram
contains a substantially larger amount of peaks and baseline
instability, then that is an indication that there is contamination
upstream of the capillary column (ie. contaminated inlet, dirty
carrier gas, etc.).
- If BOTH chromatograms contain
few peaks or very little baseline drift, it can be assumed that the
carrier gas and/or inlet are relatively clean.
- If BOTH chromatograms
contain a significant amount of noise and/or baseline drift, then
that usually is an indication that the detector or detector gases
are contaminated.
.
Back to
Top
"Low-Bleed"
Columns - Fact or Fiction?
Prof.
Walt Jennings
Cofounder, J&W Scientific Incorporated
Several manufacturers
offer "low bleed" columns. In some cases, these are merely
selected from the standard production process, but in other cases the
columns are actually "synthesized" for low bleed. In recent
years, it has been established that where functional groups (i.e.
phenyl) are inserted into the polysiloxane chain as aryl inclusions, as
opposed to being attached to the chain as pendant groups, the resultant
phase possesses increased thermal and oxidative resistance. Columns
coated with such phases emit lower levels of bleed signal and are
capable of going to higher temperatures. The increased thermal
resistance is apparent only at temperatures above ca. 300 degrees. While
some users can reap the benefits of these developments, others find
little or no improvement.. their bleed signals are still too high.
True column bleed, of
course, comes only from the column. What the user perceives as bleed is
usually the total signal reaching the detector, which is the summation
of signal from the septum
(this gives a typical silicone mass spectrum), the injector, and the
detector, all of which is usually blamed on the column.
It is good procedure to
first check the detector. Disconnect and remove the column, and place an
undrilled cap on the column attachment fitting. Activate the detector,
and note the signal at 50 degrees. Increase the oven temperature to 320
degrees, and again note the signal. On a pristine detector, the FID
signal will increase by one to two picoamps. If the increase exceeds
this level, attention should be directed to cleaning the detector,
make-up gas and hydrogen lines. Once the detector signal falls to an
acceptable level at 320 degrees, attention should be directed to the
injector. If the injector liner is visibly soiled, the injector should
be cooled, dissembled and interior cavities scrubbed with solvent and
natural bristle brushes or cotton swabs. After assembling the injector,
a "jumper tube" (one to three meters of uncoated fused silica
or steel tubing) is then used to connect the injector directly to the
detector. The injector heater should be energized, and the oven set at
320 degrees. Any increase in "bleed" signal over that observed
with the detector alone must come from the front end of the instrument,
and may originate with the septum, the carrier gas line, in-line
regulators, valves, or flow controllers.
Wrap a new septum in
aluminum foil, ensuring that one face is smooth, and install this,
smooth side down. If the signal emanating from the jumper tube is
decreased, it indicates a need for better quality septa. If the signal
is still high, materials entrained in the carrier gas may have deposited
in lines, valves, or regulators, which should be dissembled and cleaned
or replaced.
When the combined signal
from the injector and detector falls to an acceptable level (one to two
picoamps @ 320 degrees on an FID), the user is ready to install and reap
the benefits of a true low-bleed column. The bleed rate of conventional
columns is normally high enough to mask signal from the injector and
detector unless these latter are heavily contaminated. With low bleed
columns, the signal from the injector and detector assumes increased
importance. This spurious signal is not infrequently limiting, and is
usually (and incorrectly) perceived as "column bleed".
Back to
Top
|