%flg="B"%>
Frequently Asked Questions
PID/FID Calibration Curve Water Injections Guard Columns Flow on dual column assemblies Conditioning New Columns Inlet activity problems Flowmeter readings Question: Answer: Question: Answer: Question: Answer: Question: Answer: If the sample matrix is relatively "clean" (a small concentration of non-volatile compounds) and the solutes are active, the guard column should be 0.5 meter to 1 meter in length. If the sample matrix is dirty, the guard column should be longer (to collect the nonvolatile compounds). Five to ten meters help simplify system maintenance. With use a guard columns saturates and it becomes necessary to replace it. The longer guard column allows the user to simply cut off the first meter or so and reinstall it into the injector instead of replacing the entire guard column. Flow on dual column assemblies Question: Answer: Connect the columns with the Y splitter and guard column. Verify the integrity of the connection. Heat the columns to a temperature at which you can inject a detectable, unretained compound.1 I like 150°C. Note the elution time of the compound. If it elutes more than 0.1 minute apart, cut 10-15 cm from the column with the later time. Repeat the process until the nonretained compound elutes within 0.1 minute on both columns, making the column's flow rates nearly the same. Run a standard under typical run conditions until resolution criteria are met. Pick a member of a pair of compounds that is difficult to resolve on one or both columns. Raise the column oven temperature high enough so that it will elute the compound between 5 and 10 minutes, inject it and note the elution times on both columns. When installing new dual columns for the same analysis, repeat the steps in paragraph two, inject the chosen compound and adjust the head pressure until the retention time for both columns is within a percentage or two of the previously recorded times. For a list of compounds for different detectors, refer to: Rood, D. A Practical Guide to the Care, Maintenance, and Trouble Shooting of Capillary Gas Chromatography Systems; Huthig, Heidelberg, 1991. Question: Answer: When you install your column, purge it with at least three volumes of carrier gas prior to ramping it to the conditioning temperature. The total column conditioning time will depend on the type of application you're running and how much bleed is acceptable. The lower the detection limit that's needed, the longer the column will need to be conditioned. (Column bleed is closely related to the polarity and the film thickness of the stationary phase.) Polar and thick film columns bleed more and require more conditioning. For most applications, 30-60 minutes of conditioning is usually sufficient. But how can you really determine when a column is sufficiently conditioned? A flame ionization detector (FID) works best for monitoring the baseline during conditioning. Toward the end of the temperature ramp (i.e., 30-40°C below the isothermal upper temperature limit), the baseline will rise, then come down and level off, at which time you may consider the column conditioned. There are those that report detector fouling during conditioning when using other types of detectors (e.g., ECD, MS), but it's generally considered a safe practice to condition the column while connected to these detectors. One more thing: don't condition a column overnight. Column life expectancy is greatly reduced when the column is stored at high temperatures. If you're experiencing an excessive amount of bleed for more than two hours, bring the oven down to room temperature and locate the source of the problem (usually oxygen entering the column from loose fittings or a leaky septum). Baseline signals that mimic column bleed can also originate from residues present in the GC itself. One more note: if the column has not been in use for a while, a mild conditioning step may be needed to drive off contamination which may have condensed inside the column during storage. Also, there is nothing to suggest a limit to the ramp rate of the oven when conditioning a column. Question: Answer: The deactivation process entails two basic steps: a leaching step to remove metal oxides at the glass surface and a derivatization step to deactivate surface silanols. Leaching involves soaking the inlet liner in a 25% mineral acid solution (e.g., hydrochloric, nitric, and sulfuric acids, but not chromic acid), usually overnight at room temperature. This portion of the deactivation process can be shortened to several hours if the acid solution is mildly heated (65°C). The derivatization step is more involved. After leaching, the liner is heated to remove free and bound water from the surface of the glass2), and then it is derivatized with a chemical agent to deactivate the surface silanol groups The choices for derivatizing agents are numerous, and methods are just as varied. Although simple deactivation procedures exist, and are fairly effective (40-50%), the procedure is a very thorough deactivation procedure, which produces a more chemically inert liner than is commonly commercially available. This procedure is especially effective for very active compounds. For more information on the properties of glass and chemical deactivation, we recommend two books by Walt Jennings: Analytical Gas Chromatography, Academic Press, and Comparisons of Fused Silica and Other Glass Columns in Gas Chromatography, Hüthig. Also, Dean Rood's book, A Practical Guide to the Care, Maintenance, and Troubleshooting of Capillary Gas Chromatographic Systems, offers discussions on poor peak shape and activity phenomena. Question: Answer: |