Firmware Download
Agilent 1200/1100 System Firmware A.06.02/06/B.01.02/06/B.02.05 (supported revision)
CURRENT FIRMWARE A.06.02/06, B.01.06 and B.02.05 AGILENT 1100/1200 SYSTEMS
Last update November 11, 2008
Download
Problems? Ask your question
Content This set provides all firmware files for all Agilent 1100/1200 series modules
  • A.06.02/06 (for 1100/1200 modules without on-board LAN)
  • B.01.02/06 (for 1100/1200 modules with on-board LAN)
  • B.02.0x (for G4208A Instant Pilot) - for latest version see later release set
  • B.04.02 (for G1323B Control Module)

and corresponding documents (revision information and compatibility matrices).

If you need a different firmware revision, return to the overview page for 1100/1200 modules firmware and select the required set.

The firmware update tools are not available in this set. Return to the overview page for 1100/1200 modules firmware and select "Agilent 1100/1200 Systems FW Update Tools"  for more information on the tools and procedures.

The A.06.02 firmware release has been introduced with the introduction of the Agilent 1200 series in March 2006. It allows to upgrade existing 1100 series modules with firmware revision A.06.01 and below when working with the the new 1200 series.

Upgrade ALL existing 1100 modules to A.06.02/03 only if

  • a new 1200 module is added to the existing system (recommended)
  • one of the existing modules requires an upgrade due to a bug fix solved with A.06.02/03.
Read the document FWUpdate very carefully for updating to revision A.06.02 and above or downgrading to previous revisions ("1100 emulation mode").

Use only LAN/RS-232 FW Update Tool 2.6 [003] or above for firmware upgrades/downgrades (see document FWUpdate).

NOTE: See important info about compatibility below!

(latest versions, sorted by product number)

G1157A 2PS/10PT Valve (1100/1200) A.06.02 [001]  
G1158A/B 2PS/6PT Valve (1100/1200) A.06.02 [001]  
G1159A 6PS Selection Valve (1100/1200) A.06.02 [001]  
G1160A 12PS/13PT Selection Valve (1100/1200) A.06.02 [001]  
G1162A 2PS/6PT Micro Valve (1100/1200) A.06.02 [001]  
G1163A 2PS/10PT Micro Valve (1100/1200) A.06.02 [001]  
G1310A Isocratic Pump (1100/1200) A.06.04 [001]  
G1311A Quaternary Pump (1100/1200) A.06.04 [001]  
G1312A Binary Pump (1100/1200) A.06.04 [001]  
G1312B Binary Pump SL (1200) A.06.06 [001] 05/31/07
G1313A Autosampler (1100) A.06.04 [003]  
G1314A VWD (1100) A.06.04 [001]  
G1314B VWD (1200) A.06.04 [001]  
G1314C VWD SL (1200) A.06.04 [001]  
G1315A/B DAD (1100/1200) A.06.04 [001]  
G1315C DAD SL (1100/1200) B.01.06 [001] 04/27/07
G1315D DAD (1200) B.01.06 [001] 04/27/07
G1316A TCC (1100/1200) A.06.04 [001]  
G1316B TCC SL (1200) A.06.04 [001]  
G1321A FLD (1100/1200) A.06.02 [001]  
G1323B Control Module (1100/1200) B.04.02  
G1329A Thermostatted ALS (1100/1200) A.06.04 [003]  
G1329B Thermostatted ALS SL (1200) A.06.04 [003]  
G1361A Prep LC Pump (1100/1200) A.06.04 [002]  
G1362A RID (1100/1200) A.06.02 [001]  
G1364A/D Automated Fraction Collectors (1100/1200) A.06.05 [001]  
G1365A/B MWD (1100/1200) A.06.04 [001]  
G1365C MWD SL (1100/1200) B.01.06 [001] 04/27/07
G1365D MWD (1200) B.01.06 [001] 04/27/07
G1367A 1100 Well Plate Sampler (1100) A.06.03 [001]  
G1367B High Performance Autosampler (1200) A.06.03 [001]  
G1367C High Performance Autosampler SL (1200) A.06.03 [001]  
G1367D High Performance Autosampler SL+ (1200)
Backward Compatibility Release (acting as G1367C)
A.06.03 [020] 10/13/08
G1376A Capillary Pump (1100/1200) A.06.11 [005] 08/26/08
G1376A Capillary Pump (1100/1200) A.06.02 [001]  
G1377A Micro Well Plate Sampler (1100/1200) A.06.03 [001]  
G1389A Micro ALS (1100/1200) A.06.04 [003]  
G1390A Universal Interface Box (UIB) (1100/1200) A.06.02 [002]  
G2226A Nano Pump (1100/1200) A.06.02 [001]  
G2254A Automation Interface (1100/1200) A.06.02 [001]  
G2258A Dual Loop Autosampler (1100/1200) A.06.03 [001]  
G2260A Preparative Autosampler (1100/1200) A.06.04 [003]  
G4240 Chip Cube (1100/1200) A.06.05 [001]  
G4208A Instant Pilot (1200) B.02.05 [002] 06/13/07
RESIDENT FW 1100/1200 A.06.02 [001]  
RESIDENT FW for G1315C/D/G1365CD B.01.06 [001] 04/27/07
RESIDENT FW for G4208A B.02.05 [002] 06/13/07
Firmware History A.06.02/05/B.01.02/06/B.02.01/05 10/13/08 10/13/08
Compatibility Matrix A.06.02//06/B.01.02/06/B.02.01/06 06/13/07 06/13/07
Note: [nnn] build number of firmware, implemented with release A.05.01

 

COMPATIBILITY ISSUES:
This firmware has been introduced with the Agilent 1200 series of HPLC modules.

Refer to the compatibility matrix documents included (especially for the firmware for the new modules G1315D, G1365D and G1329B).

  • Firmware A.06.02/06 and B.01.0x and B.02.01 is NOT compatible with ALL previous revisions (A.06.01 and below)!
  • NO mixed systems are allowed, since they do not communicate with each other.
  • ALL modules in a system must be upgraded to A.06.02/05 PRIOR to be connected to modules with firmware A.06.02 and above.
  • Firmware A.06.02/03 is released and tested with
    • ChemStation B.02.01
    • ChemStation B.01.03 SR 1
    • Cerity A.02.2 SR2
    • Cerity A.02.03 SR1
    • Cerity A.02.03 SR2
    • EZChrom 3.1.7
    • Handheld Control Module B.04.02
    • Instant Pilot B.01.02/03
  • Firmware A.06.02/03/04/05 runs also on Agilent ChemStations B.01.03 and A.10.02 (except for the new features)
  • Control Module G1323B with firmware B.03.22 is not compatible with A.06.0x and G1315C DAD SL or G1365C MWD SL or G4240 Chip Cube. Upgrade your G1323B to new firmware release B.04.02.
  • Using the new G1323B release (B.04.02) with firmware A.05.11 and below may show incompatibilities. Therefore we recommend to install the B.04.02 only on those G1323B that are connected to systems with A.06.01 and above.


Additional Keywords :
1100, 1200, firmware
 
 
Documents
TitleSize (MB)Short Description
Firmware Information0.14This set contains all firmware files for all Agilent 1100/1200 series modules (including Instant Pilot G4208A and Control Module G1323B) and corresponding documents (revision information and compatibility matrices)
Downloads

 
 File NameFile DateFile Size
1100_1200_Firmware_44950.pdf 11/17/2009 150018
Fwupgrad\LC\1156A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1157A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1158AB_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1159A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1160A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1162A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1163A_A602_001.dlb 6/18/2007 49605
Fwupgrad\LC\1310A_A604_001.dlb 6/18/2007 642186
Fwupgrad\LC\1311A_A604_001.dlb 6/18/2007 642186
Fwupgrad\LC\1312A_A604_001.dlb 6/18/2007 642186
Fwupgrad\LC\1312B_A606_001.dlb 6/18/2007 626661
Fwupgrad\LC\1313A_A604_003.dlb 6/18/2007 686341
Fwupgrad\LC\1314AB_A604_001.dlb 6/18/2007 629342
Fwupgrad\LC\1314C_A604_001.dlb 6/18/2007 629336
Fwupgrad\LC\1315AB_A604_001.dlb 6/18/2007 718644
Fwupgrad\LC\1315C_B106_001.dlb 6/18/2007 3490874
Fwupgrad\LC\1315D_B106_001.dlb 6/18/2007 3490874
Fwupgrad\LC\1316A_A604_001.dlb 6/18/2007 601970
Fwupgrad\LC\1316B_A604_001.dlb 6/18/2007 601970
Fwupgrad\LC\1321A_A602_001.dlb 6/18/2007 669280
Fwupgrad\LC\1329A_A604_003.dlb 6/18/2007 686341
Fwupgrad\LC\1329B_A604_003.dlb 6/18/2007 686341
Fwupgrad\LC\1361A_A604_002.dlb 6/18/2007 625507
Fwupgrad\LC\1362A_A602_001.dlb 6/18/2007 683633
Fwupgrad\LC\1364A-D_A605_001.dlb 6/18/2007 718440
Fwupgrad\LC\1365AB_A604_001.dlb 6/18/2007 718644
Fwupgrad\LC\1365C_B106_001.dlb 6/18/2007 3490874
Fwupgrad\LC\1365D_B106_001.dlb 6/18/2007 3490874
Fwupgrad\LC\1367A_A603_001.dlb 6/18/2007 666090
Fwupgrad\LC\1367B_A603_001.dlb 6/18/2007 666090
Fwupgrad\LC\1367C_A603_001.dlb 6/18/2007 666084
Fwupgrad\LC\1367D_A603_020.dlb 10/13/2008 651520
Fwupgrad\LC\1376A_A602_001.dlb 6/18/2007 651369
Fwupgrad\LC\1377A_A603_001.dlb 6/18/2007 666084
Fwupgrad\LC\1389A_A604_003.dlb 6/18/2007 686341
Fwupgrad\LC\1390A_A602_002.dlb 6/18/2007 56637
Fwupgrad\LC\2226A_A602_001.dlb 6/18/2007 651369
Fwupgrad\LC\2254A_A602_001.dlb 6/18/2007 58710
Fwupgrad\LC\2258A_A603_001.dlb 6/18/2007 666084
Fwupgrad\LC\2260A_A604_003.dlb 6/18/2007 686341
Fwupgrad\LC\4208A_B205_002.dlb 6/18/2007 4869039
Fwupgrad\LC\4240A_A605_001.dlb 6/18/2007 711376
Fwupgrad\LC\config.sys 6/18/2007 101
Fwupgrad\LC\FW Compatibility-A.06.06_B.01.06.xls 11/10/2008 130560
Fwupgrad\LC\FW_A0602.pdf 11/10/2008 667751
Fwupgrad\LC\lcb402en.bin 6/18/2007 1900544
Fwupgrad\LC\res_4208A_B205_002.dlb 6/18/2007 1858721
Fwupgrad\LC\res_A602_001.dlb 6/18/2007 129294
Fwupgrad\LC\res_B106_001.dlb 6/18/2007 1707512
Fwupgrad\LC\UPDATE.EXE 6/18/2007 22772
Total Files: 51 42.4 MB