Subject: G6410A-73 G6410A-73 Supersedes None Release of Patch 2 for MassHunter Workstation QQQ B.01.04 Acquisition Service Serial Numbers: US00000000-US99999999 Release of Patch 2 for MassHunter Workstation QQQ B.01.04 Acquisition To Be Performed By: Agilent-Qualified Personnel or Customer (for Both) Parts Required: G3335-60046 MassHunter Workstation QQQ Acquisition B.01.04 Patch 2 CD-ROM Patch 2 by itself for MassHunter Workstation QQQ Acquisition B.01.04 is available at: \\wscsfs01.agilent.com\Lsca\apgsup\Software\CURRENT\ G3335AA MassHunter Acquisition for QQQ B.01.04 Build 84\Patches\Patch 2 and will be available shortly on the Agilent external website. https://www.chem.agilent.com/scripts/cag_register.asp?toURL=/scripts/cag_checkreg.asp?anch=MSQQQ Situation: On February 13, 2009 the software for the Agilent 6410 Triple Quad LC/MS went through a release for a patch version that addresses known problem summarized in the Software Status Bulletin and documented in Service Notes. For MassHunter Workstation QQQ Acquisition B.01.04 Patch 2, the following have been addressed: -- SCR# 12696 "Only 4 Scan Segments allowed when in MS2Scan Mode." Any acquisition with MS2 in Scan needs to be limited to 4 Scan Segments in MassHunter Acq. B.01.04 per Service Note G6410A-53. So for MS2Scan, Product Ion Scan, and Neutral Loss, the Scan Segments are temporarily limited to 4. -- SCR# 12717 "Dipping Quad Drivers may return nanKHz Value." Per Service Note G6410A-57 pipping Quad Drivers may return nanKHz Value with Firmware version A.04.99. New Quad Driver dipping algorithm corrects this issue with the included firmware version A.05.107. -- SCR# 13363 "Positive-Negative Swtiching in Same Time Segment Works for G6410A and G6410A-2k." G6410A and G6410A-2k QQQ LC/MS do not support Positive- Negative Switching in the Same Segment per Service Note G6410A-64. -- SCR# 14163 "Lens2DC and Lens2RF dynamic ramps can be disabled." Do not save Tune File in MS2 Scan, MS2 SIM, or MS2 Profile in Manual Tune Context from Service Note G6410A-70 is superseded. -- ALS-SL+ name correction for G1367D as reported in Service Note G6410A-63. -- Firmware A.04.105 now places the instrument into standby when certain sources are opened. It has been identified that some sources like electrospray (ESI) when opened turn off the voltages , but the instrument stays ready (green icon). Solution/Action: To install Patch 2: 1) Shutdown MassHunter Acquisition rev. B.01.04 or B.01.04 Patch 1 and shut down the Engines (or MassHunter processes). 2) Copy the file 'QQQ Acq B.01.04 Patch 02.zip' to the hard drive of the instrument control PC. 3) Uncompress 'QQQ Acq B.01.04 Patch 01.zip' to an accessible location and in the folder, locate 'QQQ Acq B.01.04 Patch 02.exe' 4) Double-click on the file 'QQQ Acq B.01.04 Patch 02.exe'. When prompted, click on OK. The patch will install. 5) Enter the instrument IP address default of 192.168.254.12 to initiate firmware update when prompted. Afterwards, power cycle the instrument or reset the smartcard to complete successful update, and click on OK for patch completion 6) If the instrument control PC has had an Installation Qualification (IQ) run on it and when the patch is complete, the user will be prompted to run an Installation Qualification Test (IQT). The test will verify that the patch is correctly installed. Additional Information: Patch 2 is a cumulative patch that includes all the updates for Patch 1. This includes: -- Support for the new G1367D 1200 LC series high performance autosampler, ALS-SL+ (minimum LC firmware A.06.12). -- Support for G3793AA MassHunter Optimizer software version B.01.04 SR-1. -- Improved QQQ firmware A.04.104 for better positive-negative switching reliability, Therefore, Service Note G6410A-63 "Release of Patch 1 for MassHunter Workstation QQQ B.01.04 Acquisition" is superseded. MassHunter Workstation B.01.04 Patch 2 will be pre-installed on the HP Workstaion bundles starting in February. Please verify firmware on QQQ instrument matches as version A.00.4.105. Install as necessary. Date: 10/29/2008 ****************************************************************************** * * * Modification Recommended * * * ****************************************************************************** * * * Standards: Labor hours: 0.00 / Travel: None * * Action Category: Immediately * * Location Category: Customer installable * * Service Inventory: See text * * Used parts: See text * * Availability: Product Support Life * * Agilent Responsible Until: 12/31/2009 * * Author/Entity: JC/2300 * * Additional Information: None * * * ****************************************************************************** Copyright 2009 Agilent Technologies Printed in U.S.A.