LMD      G4600AA-GC

Known Problem Report as of Mar 18 2008 5:10PM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software product designated above. The SSB is derived from Known Problem Reports (KPR) which result from user problems that have been classified as documentation problems or software defects. When a KPR is written, an identifying number is assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented in the SSB. User problems that have been submitted, but that have not been classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the problem has been reported already, and if there is a temporary workaround available for the problem, or if the problem has already been fixed by a new revision. If the problem is not listed in this SSB then you may wish to report it to the Response Center or to your field support representative.

To determine if your problem is documented in this SSB, first look in the "Keyword Index" section of the SSB. Under each keyword is a listing of one-line descriptions of related KPRs. If any of these sound like yours, locate the KPR # in the "Known Problem Reports" section of the SSB, and read the full KPR. The KPRs in the "Known Problem Reports" section are sorted by KPR #.

There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a brief description and a KPR #. A KPR may be associated with more than one keyword.

Known Problem Reports: This section contains KPRs, with all the available information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary


n/a
6890 Control
Checkout Test
Chromatographic Attribute
LMD Logbook
Report
Status Report

Keyword Index


Keyword: n/a

LMD A.01.03 - Must close 7890 Chemstation to Reset EMF Counters in LMD
KPR#:1314
monitoring 4 items on one instruemnt at the same time when only one reached thel
KPR#:1761
Trouble adding and removing new users
KPR#:1847
LMD UI Incomplete
KPR#:1906
We now have an active stop test button but it doesent work
KPR#:1963
Injection Based EMF counters do not Cause Not Ready State
KPR#:2000
No activity after pressing cancel or the "red X" on the sample/ vial # pop up
KPR#:2021
LMD does not disconnect from 7890 GC after test fails or is stopped by user
KPR#:2171
S/SL Pressure Decay test will not change inlet to Split - test will fail
KPR#:2172
Purge Time defaults to user entered input on the GC
KPR#:2205

Keyword: 6890 Control

Can not regain control of 6890 after stopping the GC or GC times out
KPR#:844

Keyword: Checkout Test

Injector syringe fault locks up Checkout test
KPR#:359

Keyword: Chromatographic Attribute

ChromAttrib: enlarged display loses various sections of plotted chromatograms
KPR#:681
ChromAttrib: Mis-identification possible for closely spaced peaks
KPR#:698
unable to reuse the name of a rejected standard
KPR#:1769

Keyword: LMD Logbook

7890A Firmware revision is wrong in Event Viewer Log after update
KPR#:708

Keyword: Report

Logs&Results time value can be off by ~1 min as result of truncation
KPR#:750
Status Report - Setpoints not reported as Off
KPR#:815

Keyword: Status Report

Garbage: Export Status Report to PDF File
KPR#:2158


KPR#: 1314    Product: LMD     G4600AA-GC     A.01.03

Keyword: n/a

One-line Description:
LMD A.01.03 - Must close 7890 Chemstation to Reset EMF Counters in LMD

Problem:

LMD A.01.03 requires the user to close the 7890 Chemstation in order to reset a Resource Counter. This is fixed in rev. 2.01

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 1761    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
monitoring 4 items on one instruemnt at the same time when only one reached thel

Problem:

monitoring 4 items on one instruemnt at the same time when one (the first one) reached the limit everythng work as expected except the LAAG screen report 4 advisories for that instrument This is an intermittent problem that is schedule to be investigated and fixed for 2.02

Temporary Solution:

After counters are reset the problem should disappear

Fix Information:

n/a


KPR#: 1847    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
Trouble adding and removing new users

Problem:

When attempting to remove an added user the administrator check box remains. The user name and ID go away leaving only the check box behind. If a new user is then added to LMD, that name will fill the space left behind.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 1906    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
LMD UI Incomplete

Problem:

When installing LMD to a directory OTHER than the default, Help links do not appear in the LMD UI.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 1963    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
We now have an active stop test button but it doesent work

Problem:

While running a test and pressing the stop test button, the test stops but the 6890 GC continues to run and is not released

Temporary Solution:

Turn the power off then on again To release the 6890 GC

Fix Information:

n/a


KPR#: 2000    Product: LMD     G4600AA-GC & G4800AA-LC     A.02.01

Keyword: n/a

One-line Description:
Injection Based EMF counters do not Cause Not Ready State

Problem:

for 7890s when the defined error threshold of an injection based counter is crossed and the Counter has the "Set Not Ready" option checked, the GC does not go into a Not Ready State. however Time Based counters will set not Ready.

Temporary Solution:

No workaround exists

Fix Information:

n/a


KPR#: 2021    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
No activity after pressing cancel or the "red X" on the sample/ vial # pop up

Problem:

Dialog remains open if cancel or X is clicked

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 2171    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
LMD does not disconnect from 7890 GC after test fails or is stopped by user

Problem:

LMD does not disconnect from 7890 GC after test fails or is stopped by user

Temporary Solution:

After this happens power cycle the 7890 GC and it will release

Fix Information:

n/a


KPR#: 2172    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
S/SL Pressure Decay test will not change inlet to Split - test will fail

Problem:

S/SL Pressure Decay test The inlet must be set to SPLIT or the test will fail. Executing the test will not change the inlet from Splitless to Split. The way the test is set up, it will fail if the inlet is Splitless. When the GC inlet is splitless, Septum purge remains on and the GC will never ready after the inlet is capped off.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 2205    Product: LMD     G4600AA-GC     A.02.01

Keyword: n/a

One-line Description:
Purge Time defaults to user entered input on the GC

Problem:

When running a Checkout test, LMD does not change Purge Time when the method is downloaded. The GC reads what is on the front panel. This can cause an FID Checkout Sample to fail.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 844    Product: LMD     G4600AA-GC     n/a

Keyword: 6890 Control

One-line Description:
Can not regain control of 6890 after stopping the GC or GC times out

Problem:

Lab Monitor & Diagnostic SW does not release control of the 6890 when the GC is stopped or times out. Power cycle of the GC does not fully release it. There is no way to Over-ride Lab Monitor & Diagnostic control to release the GC.

Temporary Solution:

The GC must be power cycled and the Agilent Instrument Monitor Service must be stopped and re-started from Services and Applications.

Fix Information:

n/a


KPR#: 359    Product: LMD     G4600AA-GC     n/a

Keyword: Checkout Test

One-line Description:
Injector syringe fault locks up Checkout test

Problem:

7890A FID Checkout as well as 6890 halts on step 6 "Peforming FID checkout method run" after the injector gets a syringe error. Selecting "Stop Test" from the user interface doesn't clear display or the Remote light from 7890 or 6890.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 681    Product: LMD     G4600AA-GC     n/a

Keyword: Chromatographic Attribute

One-line Description:
ChromAttrib: enlarged display loses various sections of plotted chromatograms

Problem:

When the right mouse button is dragged slowly along the amplitude axis the level of magnification of the two overlaid chromatograms is adjusted. Sections of one or both of the chromatograms, their labels and the background "grid" start mysteriously to disappear and then reappear. This behavior appears to cycle as the magnification changes so that a "missing" display can be recovered if the magnification is carefully adjusted.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 698    Product: LMD     G4600AA-GC     n/a

Keyword: Chromatographic Attribute

One-line Description:
ChromAttrib: Mis-identification possible for closely spaced peaks

Problem:

ChromAttrib: Mis-identification possible for closely spaced peaks

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 1769    Product: LMD     G4600AA-GC     A.02.01

Keyword: Chromatographic Attribute

One-line Description:
unable to reuse the name of a rejected standard

Problem:

While running chromatographic attributes to acquire a standard and using any given standard name. Rejecting the standard and leaving the CA screen disables the ability to reuse the same name. When returning to the CA screen to run another standard using the previously used name the message "standard name already in use" appears.

Temporary Solution:

If you do not navigate away from chromatographic attributes before trying to re use the name then the user can reuse the name. Or If you delete the standard name and then navigate away from the page and come back then you can reuse the name.

Fix Information:

n/a


KPR#: 708    Product: LMD     G4600AA-GC     n/a

Keyword: LMD Logbook

One-line Description:
7890A Firmware revision is wrong in Event Viewer Log after update

Problem:

The 7890A Firmware was updated to A.01.03. The firmware revision is shown as A.01.00 after the update in the Event Viewer Log.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 750    Product: LMD     G4600AA-GC     n/a

Keyword: Report

One-line Description:
Logs&Results time value can be off by ~1 min as result of truncation

Problem:

Chromatographic Attributes comparison run made at 1:50:52 shows up in table as 1:50. The time is not rounded to the nearest minute.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 815    Product: LMD     G4600AA-GC     n/a

Keyword: Report

One-line Description:
Status Report - Setpoints not reported as Off

Problem:

In the status report, the inlet pressure and flow and the detector flows are not reported as Off if they are turned off. Instead, they are reported as what they would be if they were turned on. For example, set the flow to 10 ml/min and turn it on. The status report shows the value as 10 ml/min. Turn the flow off, and the report still shows 10 ml/min. This is confusing and a problem when diagnosing a problem, as the setpoint can be shown as a relatively high value but actually be off.

Temporary Solution:

n/a

Fix Information:

n/a


KPR#: 2158    Product: LMD     G4600AA-GC & G4800AA-LC     A.02.01

Keyword: Status Report

One-line Description:
Garbage: Export Status Report to PDF File

Problem:

n/a

Temporary Solution:

n/a

Fix Information:

n/a