Specifications - 1200 Series LC/MS Systems 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1

High-performance design features
Orthogonal-flow sprayer with pneumatic nebulizer Enhances sensitivity across a wide range of LC conditions and reduces maintenance requirements.
High-capacity solvent drying system Reduces chemical noise and permits the analysis of thermally-sensitive compounds through user control of gas flow rate and gas temperature.
Dielectric platinum-plated capillary Enhances sensitivity, permits independent optimization of the spray chamber and ion optics, and allows dynamic optimization of in-source fragmentation.
Air-cooled turbomolecular pump Provides fast startup and a small footprint, and eliminates the need for cooling water. Single, split-flow turbo pump enhances performance and increases reliability.
Fraction collection capability Can monitor and trigger the collection of fractions based on mass, an analog signal input, or time. For mass-based determination, up to 16 m/z values can be monitored. Requires optional I/O Accessory Module.

Mass Range
VL m/z 10 - 1500
SL m/z 2 - 3000

Mass Accuracy
± 0.13 u within the calibrated mass range in scan mode

Mass Axis Stability
± 0.13 u over 8 hours

Scan Speed
VL 2500 u/s
SL 2500 u/s standard mode
  5250 u/s fast scan mode

SIM Sensitivity
ESI at 400 µ/min or
APCI at 1000 µ/min
Selected ion monitoring of m/z 609.3
Positive ionization

SIM Sensitivity
VL 10 pg reserpine 100:1 RMS (20:1 peak-to-peak)
SL 1 pg reserpine 100:1 RMS (20:1 peak-to-peak)

Note: When using the Multimode source in ESI mode, the same sensitivity applies.
When using the Multimode source in APCI mode, the sensitivity is 50:1 RMS
(10:1 peak-to-peak). In mixed mode operation a test mix at installation will provide
no less than 20 % of the response obtained with dedicated sources.


Scan Sensitivity
ESI at 400 µ/min or
APCI at 1000 µ/min
Scan range m/z 100 – 650
Scan speed 2500 u/s
Extracted ion at m/z 609.3
Positive ionization

Scan Sensitivity*
SL 50 pg reserpine 100:1 RMS (20:1 peak-to-peak)
*There is no scan specification for the VL model

Flow Rate
ESI Up to 1.0 ml/min water without a flow splitter
APCI Up to 1.5 ml/min water without a flow splitter

Multisignal acquisition (SL model only)
Ability to cycle through four acquisition modes on a scan-by-scan basis within a single run. Up to four analysis modes can be viewed in the software window and can be linked together to facilitate processing and analysis. Modes of operation: positive/negative polarity switching, variable fragmentor voltage selection, and concurrent SIM/scan.

Operating Conditions
Temperature range 15° to 35°C (59° to 95°F)
Relative humidity less than 95% (non-condensing)

Dimensions - Size and Weight*
Height 58 cm (23 in)
Width 74 cm (29 in)
Depth 66 cm (26 in)
Weight 63 kg (139 lb)
*Approximate maximum cabinet dimensions with G1947A APCI source attached but without external mechanical pump. Height and width slightly less with G1948A ESI source.

Descriptions and specifications supersede all previous information and are subject
to change without notice.
 
 