| Syringe size |
100 µL |
| Loop sizes for DLW and maximum pressure rating |
2 µL (standard) 1200 bar max., 5 µL, 10 µL, 20 µL 1000 bar max., 50 µL and 100 µL 600 bar max. |
| LC injector |
1 electrically actuated fast switching 6-port valve from Rheodyne (1200 bar) |
| Injection volumes |
Up to 100 µL with standard setup, up to 5 ml with additional hardware |
| HTC sample capacity |
Up to 24 shallow well-plates (96 or 384) Up to 12 deep well-plates (96 or 384), or 648 2-mL vials Up to 12 shallow well-plates (96 or 384), thermostatted Up to 6 deep well-plates (96 or 384), or 324 2-mL vials, thermostatted |
| Injection precision |
<0. 5% with 100 ul syringe and 2 ul loop, overfill 7ul |
| HTS sample capacity |
Up to 24 shallow well-plates (96 or 384) Up to 18 deep well-plates (96 or 384), or 972 2-mL vials Up to 12 deep well-plates (96 or 384), or 648 2-mL vials, thermostatted |
| Needle, syringe valve cleaning |
Dynamic load and wash (DLW) principle: two different solvents |
| Carryover |
Typically < 0.004% with DLW Method: Chlorohexidine, 0.6 mg/L dissolved in H2O : ACN = 90 : 10, 0.1% TFA Blank Solution: H2O : ACN = 90 : 10, 0.1% TFA |
| Local control |
Handheld control panel with 4 function keys, graphical LCD display |
| Electronic control |
2 RS232C ports 3 TTL inputs 1 opto-coupler input 2 relay outputs |
| Thermal control |
4-40C (settable) |
|
|