|
Hydrogen (H2) Carrier Gas Safety Guide
The use of hydrogen as a GC carrier gas is potentially dangerous.
Hydrogen (H2) is a commonly used GC carrier gas. Hydrogen, however, is
potentially explosive and has other characteristics that make
it dangerous.
- Hydrogen is combustible over a wide range of concentrations.
At atmospheric pressure, hydrogen is combustible at concentrations
from 4% to 74.2% by volume.
- Hydrogen has the highest burning velocity of any gas.
- Hydrogen has a very low ignition energy.
- Hydrogen temperature increases as the gas expands.
- Hydrogen that is allowed to expand rapidly from high pressure
can self ignite.
- Hydrogen burns with a nonluminous flame which can be invisible
under bright light.
|
|
Some dangers are general, others are unique to GC/MS operation.
Hydrogen presents a number of dangers. Some are general, others are unique
to GC or GC/MS operation. Dangers include, but are not limited
to:
- Combustion of leaking hydrogen.
- Combustion due to rapid expansion of hydrogen from a high-pressure
cylinder.
- Accumulation of hydrogen in the GC oven and subsequent combustion
(see your GC documentation and the label on the top edge of the
GC oven door).
- Accumulation of hydrogen in the mass spectrometer and subsequent
combustion.
In a mass spectrometer, hydrogen can accumulate several ways
All users should be aware of the mechanisms by which hydrogen can
accumulate. They should also know what precautions to take if
they know, or suspect, that hydrogen has accumulated.
| Mechanism |
Results |
| Mass spectrometer off |
A mass spectrometer can be shut down deliberately. It can also be shut down accidentally by an internal or external failure. A mass spectrometer shutdown does not shut off the flow of carrier gas. As a result, hydrogen may slowly accumulate in the mass spectrometer. |
| Mass spectrometer automated isolation valves closed |
Some mass spectrometers are equipped with automated diffusion pump isolation valves. In these instruments, deliberate operator action or a number of failures can cause the isolation valves to close. Isolation valve closure does not shut off the flow of carrier gas. As a result, hydrogen may slowly accumulate in the mass spectrometer.
|
| Mass spectrometer manual isolation valves closed |
Some mass spectrometers are equipped with manual diffusion pump isolation valves. In these instruments, the operator can close the isolation valves. Closing the isolation valves does not shut off the flow of carrier gas. As a result, hydrogen may slowly accumulate in the mass spectrometer.
|
| GC off |
A GC can be shut down deliberately. It can also be shut down accidentally by an internal or external failure. Different GCs react in different ways.
If an HP 5890 GC equipped with Electronic Pressure Control (EPC) is shut off, the EPC does not stop the flow of carrier gas. It increases the flow of carrier gas to its maximum. This flow may be more than some mass spectrometers can pump away, resulting in the accumulation of hydrogen in the mass spectrometer. If the mass spectrometer is shut off at the same time, the accumulation can be fairly rapid.
|
| Power failure |
If the power fails, both the GC and mass spectrometer shut down. The carrier gas, however, is not necessarily shut down. As described previously, in some GCs a power failure may cause the carrier gas flow to be set to maximum. As a result, hydrogen may accumulate in the mass spectrometer.
|
|
- Install an automatic shutoff valve (such as the HP 19246S
Gas Saver for the HP 5890 GC) that will shut off hydrogen
flow in the event of a power failure.
General laboratory precautions
- Avoid leaks in the carrier gas lines. Use leak-checking equipment
to periodically check for hydrogen leaks.
- Eliminate from your laboratory as many ignition sources as
possible (open flames, devices that can spark, sources of static
electricity, etc.).
- Do not allow hydrogen from a high pressure cylinder to vent
directly to atmosphere (danger of self ignition).
- Use a hydrogen generator instead of bottled hydrogen.
Operating precautions
- Turn off the hydrogen at its source any time you shut down
the GC or MS.
- Turn off the hydrogen at its source any time you vent the
mass spectrometer
(do not heat the capillary column without carrier gas flow).
- Turn off the hydrogen at its source any time isolation valves
in a mass spectrometer are closed (do not heat the capillary column
without carrier gas flow).
- Turn off the hydrogen at its source if a power failure occurs.
- If a power failure occurs while the GC/MS system is unattended,
even if the system has restarted by itself:
- Immediately turn off the hydrogen at its source.
- Turn off the GC.
- Turn off the mass spectrometer and allow it to cool for 1
hour.
- Eliminate all potential sources of ignition in the room.
- Open the vacuum manifold of the mass spectrometer to atmosphere
(remove the manifold window or analyzer or open the side plate;
do not just open the vent valve).
- Wait at least 10 minutes to allow any hydrogen to dissipate.
- Start up the GC and mass spectrometer as normal.
- Never turn on an ion or triode pressure gauge until the mass
spectrometer has completed its pumpdown cycle.
|